首页> 美国卫生研究院文献>other >HVAD Waveform Analysis as a Noninvasive Marker of Pulmonary Capillary Wedge Pressure: A First Step Toward the Development of a Smart Left Ventricular Assist Device Pump
【2h】

HVAD Waveform Analysis as a Noninvasive Marker of Pulmonary Capillary Wedge Pressure: A First Step Toward the Development of a Smart Left Ventricular Assist Device Pump

机译:HVAD波形分析作为肺毛细血管楔压的无创标记:迈向智能左心室辅助设备泵发展的第一步

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

Flow waveforms are an important feature of the HVAD left ventricular assist device (LVAD) that provides information about HVAD function and patient hemodynamics. We assessed the properties of one specific aspect of the waveform, the slope of the ventricular filling phase (VFP), and its correlation with pulmonary capillary wedge pressure (PCWP). A total of 101 screenshots from the HVAD monitor and simultaneous hemodynamic measurements were obtained simultaneously during sequential stages of invasive hemodynamic ramp studies. Each screenshot was digitized (IGOR Pro, WaveMetrics Inc., Oswego, OR) and properties of the flow waveforms including instantaneous flow and rate of change of flow were analyzed. Ventricular filling phase slope (VFPS) was calculated for each screenshot and correlated to PCWP. Ventricular filling phase slope was significantly higher in patients with PCWP ≥ 18 mm Hg than in patients with PCWP < 18 mm Hg [6.25 (5.84–7.37) L/min/s vs. 3.27 (2.00–4.69) L/min/s, p ≤ 0.0001]. A VFPS threshold of 5.8 L/min/s predicted a PCWP ≥ 18 mm Hg with a sensitivity of 87% and specificity of 95% (AUC 0.95). Ventricular filling phase slope of the HVAD flow waveform is a novel noninvasive parameter that correlates with PCWP and can discriminate elevated versus normal or low PCWP. Automated reporting of this parameter may help clinical assessment and management of patients supported by an HVAD and may serve as the basis of a smart LVAD pump that can adapt in response to changes in a patient’s physiology.
机译:流量波形是HVAD左心室辅助设备(LVAD)的重要功能,可提供有关HVAD功能和患者血液动力学的信息。我们评估了波形的一个特定方面的属性,心室充盈期(VFP)的斜率及其与肺毛细血管楔压(PCWP)的相关性。在有创血流动力学斜坡研究的连续阶段中,同时获得了HVAD监视器和同时进行的血流动力学测量的101张屏幕截图。每个屏幕截图均被数字化(IGOR Pro,WaveMetrics Inc.,俄勒冈州奥斯威戈),并分析了包括瞬时流量和流量变化率在内的流量波形特性。计算每个屏幕截图的心室充盈相位斜率(VFPS),并将其与PCWP相关。 PCWP≥18 mm Hg的患者的心室充盈相位斜率显着高于PCWP <18 mm Hg的患者[6.25(5.84-7.37)L / min / s与3.27(2.00-4.69)L / min / s, p≤0.0001]。 VFPS阈值为5.8 L / min / s时,预测PCWP≥18 mm Hg,灵敏度为87%,特异性为95%(AUC 0.95)。 HVAD流量波形的心室充盈相位斜率是一种新颖的非侵入性参数,与PCWP相关并且可以区分PCWP升高还是正常或低。该参数的自动报告可以帮助HVAD支持的患者进行临床评估和管理,并且可以作为智能LVAD泵的基础,该泵可以适应患者的生理变化。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号