Preliminary Clinical Experience with a Combined Automated Breast Ultrasound and Digital Breast Tomosynthesis System

机译:结合自动乳房超声和数字乳房断层合成系统的初步临床经验

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。
获取外文期刊封面目录资料

摘要

We analyze the performance of a mammographically-configured, automated breast ultrasound (McABUS) scanner combined with a digital breast tomosynthesis (DBT) system. The GE Invenia ultrasound system was modified for integration with GE DBT systems. The ultrasound and DBT imaging were performed in the same mammographic compression. Our small preliminary study included thirteen cases, six of which contained invasive cancers. From analysis of these cases, current limitations and corresponding potential improvements of the system were determined. A registration analysis was performed to compare the ease of McABUS to DBT registration for this system with that of two systems designed previously. It was observed that, in comparison to data from an earlier study, the McABUS to DBT registration alignment errors for both this system and for a previously built combined system were smaller than for a previously built standalone McABUS system.
机译:我们分析了结合了乳腺X线断层合成(DBT)系统的乳腺X射线摄影配置的自动乳房超声(McABUS)扫描仪的性能。 GE Invenia超声系统经过修改,可以与GE DBT系统集成。超声和DBT成像在相同的乳房X线照片压缩下进行。我们的小型初步研究包括13个病例,其中6个包含浸润性癌症。通过对这些情况的分析,确定了系统的当前限制和相应的潜在改进。进行了注册分析,以比较该系统与先前设计的两个系统的McABUS到DBT注册的难易程度。据观察,与早期研究的数据相比,该系统和先前构建的组合系统的McABUS到DBT对准误差均小于先前独立的McABUS系统。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号