We tested the ability of a single Compton camera (CC) to produce 3-dimensional (3D) images of prompt gammas (PGs) emitted during the irradiation of a tissue-equivalent plastic phantom with proton pencil beams for clinical doses delivered at clinical dose rates. PG measurements were made with a small prototype CC placed at three different locations along the proton beam path. We evaluated the ability of the CC to produce images at each location for two clinical scenarios: 1) the delivery of a single 2 -Gy pencil beam from a hypo-fractionated treatment [~9 × 108 protons], and 2) a single pencil beam from a standard treatment [~1 × 108 protons]. Additionally, the data measured at each location were combined to simulate measurements with a larger scale, clinical CC and its ability to image shifts in the Bragg Peak (BP) range for both clinical scenarios. With our prototype CC, the location of the distal end of the BP could be seen with the CC placed up to 4 cm proximal or distal to the BP distal falloff. Using the data from the simulated full scale clinical CC, 3D images of the PG emission were produced with the delivery of as few as 1 × 108 protons, and shifts in the proton beam range as small as 2 mm could be detected for delivery of a 2 Gy spot. From these results we conclude that 3D PG imaging for proton range verification under clinical beam delivery conditions is possible with a single CC.
机译:我们测试了单个Compton相机(CC)产生质子笔形束对相当于组织的塑料体模进行辐照时产生的即时伽马(PG)的3维(3D)图像的能力,以临床剂量率提供的临床剂量。 PG测量是通过将小型原型CC放置在沿质子束路径的三个不同位置进行的。我们评估了CC在两种临床情况下在每个位置产生图像的能力:1)从超分割治疗[〜9×10 8 质子传递单根2 -Gy铅笔束],以及2)来自标准处理[〜1×10 8 质子]的单个笔形光束。此外,将在每个位置测量的数据组合起来,以模拟更大规模的测量,临床CC及其在两种临床情况下在布拉格峰(BP)范围内成像位移的能力。使用我们的原型CC,可以将CC放置在BP远端衰减的近端或远端4 cm处,从而看到BP远端的位置。利用来自模拟的全尺寸临床CC的数据,生成了PG发射的3D图像,其中传递了少至1×10 8 质子,并且质子束范围的偏移小至2可以检测到mm来传送2 Gy点。根据这些结果,我们得出结论,使用单个CC可以在临床束传输条件下进行质子范围验证的3D PG成像。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号