Reliability analysis of a sensitive and independent stabilometry parameter set

机译:敏感且独立的稳定度参数集的可靠性分析

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

Recent studies have suggested reduced independent and sensitive parameter sets for stabilometry measurements based on correlation and variance analyses. However, the reliability of these recommended parameter sets has not been studied in the literature or not in every stance type used in stabilometry assessments, for example, single leg stances. The goal of this study is to evaluate the test-retest reliability of different time-based and frequency-based parameters that are calculated from the center of pressure (CoP) during bipedal and single leg stance for 30- and 60-second measurement intervals. Thirty healthy subjects performed repeated standing trials in a bipedal stance with eyes open and eyes closed conditions and in a single leg stance with eyes open for 60 seconds. A force distribution measuring plate was used to record the CoP. The reliability of the CoP parameters was characterized by using the intraclass correlation coefficient (ICC), standard error of measurement (SEM), minimal detectable change (MDC), coefficient of variation (CV) and CV compliance rate (CVCR). Based on the ICC, SEM and MDC results, many parameters yielded fair to good reliability values, while the CoP path length yielded the highest reliability (smallest ICC > 0.67 (0.54–0.79), largest SEM% = 19.2%). Usually, frequency type parameters and extreme value parameters yielded poor reliability values. There were differences in the reliability of the maximum CoP velocity (better with 30 seconds) and mean power frequency (better with 60 seconds) parameters between the different sampling intervals.
机译:最近的研究表明,基于相关性和方差分析,可以减少用于稳定度测量的独立和敏感参数集。但是,这些推荐参数集的可靠性尚未在文献中进行研究,也未在稳定度评估中使用的每种姿势类型(例如单腿姿势)中都未进行过研究。这项研究的目的是评估在30秒和60秒的测量间隔内从双足和单腿站立期间的压力中心(CoP)计算出的基于时间和基于频率的不同参数的重测可靠性。 30名健康受试者在双脚姿势下睁开和闭眼,在单腿姿势下睁开60秒,进行了重复站立试验。力分布测量板用于记录CoP。 CoP参数的可靠性通过使用组内相关系数(ICC),标准测量误差(SEM),最小可检测变化(MDC),变异系数(CV)和CV依从率(CVCR)来表征。根据ICC,SEM和MDC结果,许多参数得出的可靠性值基本达到良好,而CoP路径长度产生了最高的可靠性(最小的ICC> 0.67(0.54-0.79),最大的SEM%= 19.2%)。通常,频率类型参数和极值参数会产生较差的可靠性值。不同采样间隔之间最大CoP速度(更好的是30秒)和平均功率频率(更好的是60秒)参数的可靠性存在差异。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号