Thallium bromide (TlBr) is a promising semiconductor detector material for positron emission tomography (PET) because it can offer very good energy resolution and 3-D segmentation capabilities, and it also provides detection efficiency surpassing that of commonly used scintillators. Energy, timing, and spatial resolution were measured for thin (<1 mm) TlBr detectors. The energy and timing resolution were measured simultaneously for the same planar 0.87 mm-thick TlBr device. An energy resolution of (6.4±1.3)% at 511 keV was achieved at −400 V bias voltage and at room temperature. A timing resolution of (27.8±4.1) ns FWHM was achieved for the same operating conditions when appropriate energy gating was applied. The intrinsic spatial resolution was measured to be 0.9 mm FWHM for a TlBr detector with metallic strip contacts of 0.5 mm pitch. As material properties improve, higher bias voltage should improve timing performance. A stack of thin detectors with finely segmented readout can create a modular detector with excellent energy and spatial resolution for PET applications.
展开▼
机译:溴化((TlBr)是一种有前途的正电子发射断层扫描(PET)半导体检测器材料,因为它可以提供非常好的能量分辨率和3-D分割功能,并且其检测效率也超过了常用的闪烁体。测量了薄型(<1 mm)TlBr检测器的能量,时间和空间分辨率。对于同一个0.87 mm厚的平面TlBr器件,同时测量了能量和时序分辨率。在-400 V偏置电压和室温下,在511 keV下实现了(6.4±1.3)%的能量分辨率。当应用适当的能量门控时,对于相同的工作条件,实现了(27.8±4.1)ns FWHM的定时分辨率。对于具有0.5mm间距的金属条触点的TlBr检测器,固有空间分辨率测得为0.9 mm FWHM。随着材料性能的提高,较高的偏置电压将改善时序性能。带有细段读数的一堆薄检测器可以为PET应用创建具有出色能量和空间分辨率的模块化检测器。