Laboratory validation and field usability assessment of a point-of-care test for serum bilirubin levels in neonates in a tropical setting

机译:在热带环境中对新生儿血清胆红素水平进行即时检验的实验室验证和现场可用性评估

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

>Background: Screening and monitoring serum bilirubin (SBR) in neonates is crucial to prevent neonatal hyperbilirubinemia (NH)-associated morbidity and mortality worldwide. A lack of resources is often a barrier for measuring SBR in developing countries. Reliable, cost-effective, easy to use point-of-care (POC) SBR tests are needed. This study aimed to evaluate the technical accuracy and usability of the Bilistick System (BS), a new bilirubin POC test, in a tropical setting. >Methods: This was a mixed-methods study, including laboratory validation of the BS, direct observation of technical procedures as performed by the midwives and midwives’ assessment of the device’s easiness of use through focus group discussions (FGD) and a self-administered questionnaire. The study was conducted in a field clinic of the Shoklo Malaria Research Unit along the Thailand-Myanmar border between January and December 2017. >Results: A total of 173 samples were tested at a median age of 4 days. BS generated an error message—providing no SBR readout—in 48.6% of the tests performed. For the tests that yielded a result, the correlation coefficient (95% CI) between BS and routine laboratory bilirubinometer SBR was 0.87 (0.77-0.93). The accuracy decreased with increasing haematocrit and at higher humidity (≥75%). Direct observation of the operators using the device and analysis of the focus group discussions and questionnaires indicated that the BS was considered easy to use and required limited training. >Conclusions: This evaluation showed that the BS, in its current formulation, does not provide reliable results for measuring SBR in a tropical, low-resource setting  but has acceptable usability features.
机译:>背景:筛查和监测新生儿的血清胆红素(SBR)对于预防全世界与新生儿高胆红素血症(NH)相关的发病率和死亡率至关重要。缺乏资源通常是衡量发展中国家SBR的障碍。需要可靠,具有成本效益,易于使用的即时护理(POC)SBR测试。这项研究旨在评估热带环境中胆红素POC测试的新技术Bbilistick System(BS)的技术准确性和可用性。 >方法:这是一项混合方法研究,包括对BS进行实验室验证,直接观察助产士执行的技术程序以及助产士通过焦点小组讨论来评估设备的易用性(FGD) )和一份自行管理的问卷。该研究于2017年1月至12月之间在泰国-缅甸边境的Shoklo疟疾研究部门的一家野外诊所进行。>结果:总共对173个样本进行了测试,平均年龄为4天。 BS在执行的48.6%的测试中生成了一条错误消息(未提供SBR读数)。对于产生结果的测试,BS与常规实验室胆红素计SBR之间的相关系数(95%CI)为0.87(0.77-0.93)。随着血细胞比容的增加和湿度的增加(≥75%),准确性降低。对使用该设备的操作员进行直接观察以及对焦点小组讨论和调查表的分析表明,BS被认为易于使用且需要有限的培训。 >结论:该评估表明,按照目前的公式,BS无法在热带,低资源环境中提供可靠的SBR测量结果,但具有可接受的可用性功能。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号