Effect of skull shape approximations in Gamma Knife dose calculations

机译:头骨形状近似值对伽玛刀剂量计算的影响

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

Doses in Leksell GammaPlan (Elekta, Stockholm, Sweden) are calculated for each shot by summing the contribution from the 201 radiation beams emitted by the Gamma Knife (Elekta), weighted for attenuation in the tissue traversed. The patient's head is modeled based on 24 skull measurements, from which the depth to the calculation point is determined for each beam. The limited number of measurement points associated with this approach can result in substantial discrepancies between the skull model used for dose calculations and the actual skull contour from computed tomography or magnetic resonance data. A review of 24 patients found that differences between the actual and approximated skull shape gave treatment time errors as large as 4.1%, although in most instances, the errors were less than 1%. The conclusion was that the Leksell GammaPlan head model provides a quick and convenient approach for specifying the shape of the patient's head in all but extreme cases, where discrepancies as large as ~5% can result.PACS number: 87.53.Ly
机译:通过将Gamma刀(Elekta)发出的201束辐射束的总和相加,计算得出的Leksell GammaPlan(Elekta,斯德哥尔摩,瑞典)中的剂量,并对经过的组织进行衰减加权。基于24个颅骨测量值对患者的头部进行建模,从中确定每个射线束到计算点的深度。与该方法相关的有限数量的测量点可能导致用于剂量计算的颅骨模型与来自计算机断层扫描或磁共振数据的实际颅骨轮廓之间存在实质性差异。对24位患者的审查发现,实际和近似头骨形状之间的差异导致治疗时间误差高达4.1%,尽管在大多数情况下,误差小于1%。结论是Leksell GammaPlan头部模型提供了一种快速便捷的方法来指定患者头部的形状,但极端情况除外,极端情况可能会导致高达5%的差异.PACS编号:87.53.Ly

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号