One of the applications of MatriXX (IBA Dosimetry) is experimental verification of dose for IMRT, VMAT, and tomotherapy. For cumulative plan verification, dose is delivered for all the treatment gantry angles to a stationary detector. Experimental calibration of MatriXX detector recommended by the manufacturer involves only AP calibration fields and does not address angular dependency of MatriXX. Angular dependency may introduce dose bias in cumulative plan verification if not corrected. For this reason, we characterized angular dependency of MatriXX and developed a method for its calibration. We found relatively large discrepancies in responses to posterior vs. anterior fields for four MatriXX (Evolution series) detectors (up to 11%), and relatively large variability of responses as a function of gantry angle in the gantry angle ranges of 91°–110° and 269°–260°. With our calibration method, the bias due to angular dependency is effectively removed in experimental verification of IMRT and VMAT plans.PACS number: 87.56Fc
机译:MatriXX(IBA剂量测定)的应用之一是IMRT,VMAT和tomotherapy剂量的实验验证。为了进行累积计划验证,将所有治疗台角度的剂量传送到固定检测器。制造商推荐的MatriXX检测器的实验校准仅涉及AP校准字段,并且不涉及MatriXX的角度依赖性。如果不纠正,角度依赖性可能会在累积计划验证中引入剂量偏差。因此,我们表征了MatriXX的角度依赖性,并开发了一种校准方法。我们发现四个MatriXX(Evolution系列)检测器对后场和前场的响应差异较大(高达11%),并且在91°–110的龙门角度范围内,响应随龙门角度的变化相对较大°和269°–260°。使用我们的校准方法,在IMRT和VMAT计划的实验验证中有效消除了因角度依赖性而产生的偏差.PACS编号:87.56Fc

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号