首页> 美国卫生研究院文献>Journal of Applied Clinical Medical Physics >Monte Carlo investigation of energy response of various detector materials in I125 and Yb169 brachytherapy dosimetry
【2h】

Monte Carlo investigation of energy response of various detector materials in I125 and Yb169 brachytherapy dosimetry

机译:蒙特卡洛研究I125和Yb169近距离放射剂量学中各种检测器材料的能量响应

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

Relative absorbed‐dose energy response correction R for different detector materials in water, PMMA and polystyrene phantoms are calculated using Monte Carlo‐based EGSnrc code system for I125 and Yb169 brachytherapy sources. The values of R obtained for I125 source are 1.41, 0.92, 3.97, 0.47, 8.32 and 1.10, respectively, for detector materials LiF, Li2B4O7,  Al2O3, diamond, silicon diode and air. These values are insensitive to source‐to‐detector distance and phantom material. For Yb169 source, R is sensitive to source‐to‐detector distance for detector materials other than air and Li2B4O7. For silicon, R increases from 3 to 4.23 when depth in water is increased from 0.5 cm to 15 cm. For Yb169 source, the values of R obtained for air and Li2B4O7 in PMMA and polystyrene phantoms are comparable to that obtained in water. However, LiF, Si and Al2O3 show enhanced response and diamond shows decreased response in PMMA and polystyrene phantoms than in water.PACS number: 87.53.Jw
机译:使用基于蒙特卡洛的EGSnrc代码系统针对 I 125 Yb 169 近距离放射治疗来源。为 获得的R值 125 探测器材料LiF,Li2B4O7,Al2O3,金刚石,硅二极管和空气的来源分别为1.41、0.92、3.97、0.47、8.32和1.10。这些值对源到检测器的距离和幻像材料不敏感。对于 Yb < / mtext> 169 源,R对除空气和Li2B4O7以外的其他探测器材料从源到探测器的距离。对于硅,当水中的深度从0.5厘米增加到15厘米时,R从3增加到4.23。对于 Yb < / mtext> 169 源,R的值PMMA和聚苯乙烯模型中的空气和Li2B4O7所获得的结果与水中的结果相当。然而,与水相比,PMMA和聚苯乙烯模型中的LiF,Si和Al2O3表现出增强的响应,而钻石显示出下降的响应.PACS数:87.53.Jw

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号