Temperature rise during removal of fractured components out of the implant body: an in vitro study comparing two ultrasonic devices and five implant types

机译:从植入物主体中取出破裂的部件过程中的温度升高:一项体外研究比较了两种超声设备和五种植入物类型

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

BackgroundUltrasonic instrumentation under magnification may facilitate mobilization of screw remnants but may induce heat trauma to surrounding bone. An increase of 5°C is considered detrimental to osseointegration. The objective of this investigation was to examine the rise in temperature of the outer implant body after 30 s of ultrasonic instrumentation to the inner part, in relation to implant type, type of ultrasonic equipment, and the use of coolants in vitro.
机译:背景技术超声仪器在放大作用下可促进螺钉残余物的动员,但可能对周围骨骼产生热损伤。升高5°C被认为对骨整合有害。这项研究的目的是检查与植入物的类型,超声设备的类型以及体外冷却剂的使用有关的超声仪器到内部30秒钟后植入物外部的温度升高情况。