首页> 美国卫生研究院文献>Journal of Applied Clinical Medical Physics >Utilization of optical tracking to assess efficacy of intracranial immobilization techniques in proton therapy
【2h】

Utilization of optical tracking to assess efficacy of intracranial immobilization techniques in proton therapy

机译:利用光学跟踪评估颅内固定技术在质子治疗中的功效

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

We present a quantitative methodology to measure head interfraction movements within intracranial masks of commercial immobilization devices used for proton radiotherapy. A three‐points tracking (3PtTrack) method was developed to measure the mask location for each treatment field over an average of 10 fractions for seven patients. Five patients were treated in supine with the Qfix Base‐of‐Skull (BoS) headframe, and two patients were treated in prone with the CIVCO Uni‐frame baseplate. Patients were first localized by an in‐room, image‐guidance (IG) system, and then the mask location was measured using the 3PtTrack method. Measured mask displacements from initial location at the first fraction are considered equivalent to the head interfraction movement within the mask. The trends of head movements and couch displacements and rotation were analyzed in three major directions. The accuracy of 3PtTrack method was shown to be within 1.0 mm based on daily measurements of a QA device after localization by the IG system for a period of three months. For seven patients, mean values of standard deviation (SD) in anterior–posterior, lateral, and superior–inferior directions were 1.1 mm, 1.4 mm, and 1.6 mm for head movements, and were 1.4 mm, 1.8 mm, and 3.4 mm for couch displacements. The mean SD values of couch rotations were 1.1°, 0.9°, and 1.1° for yaw, pitch, and roll, respectively. The overall patterns of head movements and couch displacements were similar for patients treated in either supine or prone, with larger deviations in the superior–inferior (SI) direction. A suboptimal mask fixation to the frame of the mask to the H&N frame is likely the cause for the observed larger head movements and couch displacements in the SI direction compared to other directions. The optical‐tracking methodology provided a quantitative assessment of the magnitude of head motion.PACS number: 87.55.km
机译:我们提出了一种定量的方法来测量用于质子放疗的商业固定设备的颅内面罩内的头间质运动。开发了一种三点跟踪(3PtTrack)方法来测量每个治疗场的面罩位置,平均七名患者需要10次。 5例患者使用Qfix颅底(BoS)头枕仰卧位治疗,2例患者使用CIVCO Uni-frame底板俯卧位治疗。首先通过室内图像引导(IG)系统对患者进行定位,然后使用3PtTrack方法测量面罩的位置。从第一位置的初始位置测量的遮罩位移被认为等效于遮罩内的头部屈光运动。从三个主要方向分析了头部运动,卧榻移位和旋转的趋势。根据IG系统定位三个月后对QA设备进行的每日测量,显示3PtTrack方法的精度在1.0毫米以内。对于7例患者,前后移动的标准偏差(SD)的平均值分别为:头部运动为1.1 mm,1.4 mm和1.6 mm,头部运动为1.4 mm,1.8 mm和3.4 mm沙发位移。横摇,俯仰和横摇的沙发床旋转的平均SD值分别为1.1°,0.9°和1.1°。仰卧位或俯卧位患者的头部运动和卧榻移位的总体模式相似,上下(SI)方向偏差较大。与其他方向相比,将面罩框架固定到H&N框架的面罩固定效果欠佳,很可能是导致在SI方向观察到较大的头部移动和卧榻位移的原因。光学跟踪方法提供了对头部运动幅度的定量评估.PACS数:87.55.km

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号