Investigation of optimal display size for viewing T1‐weighted MR images of the brain using a digital contrast‐detail phantom

机译:研究使用数字对比度细节体模查看大脑T1加权MR图像的最佳显示器尺寸

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

We clarified the relationship between the display size of MRI images and observer performance using a digital contrast‐detail (d‐CD) phantom. The d‐CD phantom was developed using Microsoft Visual Basic 2010 Express. It had a 512 × 512 matrix in size and a total of 100 holes, whose diameter increased stepwise from 4 to 40 pixels with a 4‐pixel interval in the vertical direction; the contrast varied stepwise in the horizontal direction. The digital driving level (DDL) of the background, the width of the DDL, and the contrast were adjustable. These parameters were determined on the basis of the actual T1‐weighted magnetic resonance (MR) images of the brain. In this study, the DDL, width, and contrast were set to 85, 20, and 1, respectively. The observer performance study was performed for three different display sizes (30 cm × 30 cm as the enlarged size, 16 cm × 16 cm as the original size, and 10 cm × 10 cm as the reduced size) using a 2‐megapixel color liquid crystal display monitor, and it was analyzed using Friedman and Wilcoxon statistical tests. The observer performances for the original display (p  0.01) and the reduced display sizes (p  0.01) were superior to that observed for the enlarged size, whereas there was no significant difference between the original display and reduced display sizes (p = 0.31). Evaluation with the digital phantom simulating MR imaging also revealed that the original and reduced display sizes were superior to the enlarged display size in observer performance. The d‐CD phantom enables a short‐term evaluation of observer performance and is useful in analyzing relationship between display size and observer performance.PACS number: 87.57.‐s
机译:我们使用数字对比度细节(d-CD)体模阐明了MRI图像的显示尺寸与观察者性能之间的关系。 d-CD幻影是使用Microsoft Visual Basic 2010 Express开发的。它的尺寸为512×512矩阵,共有100个孔,其直径从4像素逐步增加到40像素,在垂直方向上间隔为4像素;对比度沿水平方向逐步变化。背景的数字驱动级别(DDL),DDL的宽度和对比度是可调的。这些参数是根据实际的大脑T1加权磁共振(MR)图像确定的。在这项研究中,DDL,宽度和对比度分别设置为85、20和1。使用2兆像素彩色液体对三种不同的显示尺寸(30厘米×30厘米作为放大尺寸,16厘米×16厘米作为原始尺寸,10厘米×10厘米作为缩小尺寸)进行观察者性能研究液晶显示器,并使用Friedman和Wilcoxon统计测试对其进行分析。原始显示(p <0.01)和缩小显示尺寸(p <0.01)的观察者性能优于放大显示的观察者性能,而原始显示和缩小显示尺寸之间的观察者性能无显着差异(p = 0.31) )。用数字体模模拟MR成像进行的评估还显示,原始显示尺寸和缩小的显示尺寸在观察者性能方面均优于扩大的显示尺寸。 d-CD幻影可以对观察者的表现进行短期评估,对于分析显示器尺寸和观察者的表现之间的关系非常有用.PACS编号:87.57.-s

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号