Effect of electron contamination on in vivo dosimetry for lung block shielding during TBI

机译:电子污染对TBI期间肺部阻滞屏蔽体内剂量的影响

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

Our institution performs in vivo verification measurement for each of our total body irradiation (TBI) patients with optically stimulated luminescent dosimeters (OSLD). The lung block verification measurements were commonly higher than expected. The aim of this work is to understand this discrepancy and improve the accuracy of these lung block verification measurements. Initially, the thickness of the lung block was increased to provide adequate lung sparing. Further tests revealed the increase was due to electron contamination dose emanating from the lung block. The thickness of the bolus material covering the OSLD behind the lung block was increased to offset the electron contamination. In addition, the distance from the lung block to the dosimeter was evaluated for its effect on the OSLD reading and found to be clinically insignificant over the range of variability in our clinic. The results show that the improved TBI treatment technique provides for better accuracy of measured dose in vivo and consistency of patient setup.PACS number(s): 87.53.Bn, 87.53.Kn, 87.55.N‐, 87.55.Qr
机译:我们的机构使用光学刺激的发光剂量计(OSLD)对我们的每个全身照射(TBI)患者进行体内验证测量。肺部阻滞验证测量值通常高于预期值。这项工作的目的是了解这种差异并提高这些肺部阻滞验证测量的准确性。最初,增加了肺部阻滞的厚度以提供足够的肺部保留。进一步的测试表明,增加的原因是由于肺部阻滞发出电子污染剂量。覆盖肺阻塞后OSLD的推注材料厚度增加,以抵消电子污染。此外,评估了从肺阻塞到剂量计的距离对OSLD读数的影响,发现在我们诊所的变异范围内,在临床上无意义。结果表明,改进的TBI治疗技术可提供更高的体内测量剂量准确性和患者设置的一致性.PACS编号:87.53.Bn,87.53.Kn,87.55.N-,87.55.Qr

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号