Assessment of the accuracy and stability of frameless gamma knife radiosurgery

机译:评估无框伽马刀放射外科手术的准确性和稳定性

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

The aim of this study was to assess the accuracy and stability of frameless gamma knife radiosurgery (GKRS). The accuracies of the radiation isocenter and patient couch movement were evaluated by film dosimetry with a half‐year cycle. Radiation isocenter assessment with a diode detector and cone‐beam computed tomography (CBCT) image accuracy tests were performed daily with a vendor‐provided tool for one and a half years after installation. CBCT image quality was examined twice a month with a phantom. The accuracy of image coregistration using CBCT images was studied using magnetic resonance (MR) and computed tomography (CT) images of another phantom. The overall positional accuracy was measured in whole procedure tests using film dosimetry with an anthropomorphic phantom. The positional errors of the radiation isocenter at the center and at an extreme position were both less than 0.1 mm. The three‐dimensional deviation of the CBCT coordinate system was stable for one and a half years (mean 0.04 ± 0.02 mm). Image coregistration revealed a difference of 0.2 ± 0.1 mm between CT and CBCT images and a deviation of 0.4 ± 0.2 mm between MR and CBCT images. The whole procedure test of the positional accuracy of the mask‐based irradiation revealed an accuracy of 0.5 ± 0.6 mm. The radiation isocenter accuracy, patient couch movement accuracy, and Gamma Knife Icon CBCT accuracy were all approximately 0.1 mm and were stable for one and a half years. The coordinate system assigned to MR images through coregistration was more accurate than the system defined by fiducial markers. Possible patient motion during irradiation should be considered when evaluating the overall accuracy of frameless GKRS.
机译:这项研究的目的是评估无框伽马刀放射外科手术(GKRS)的准确性和稳定性。放射线等中心线和患者卧榻运动的准确性通过胶片剂量法以半年周期进行评估。在安装后的一年半中,每天使用供应商提供的工具通过二极管检测器和锥束计算机断层扫描(CBCT)图像准确性测试进行辐射等中心线评估。每月用幻像对CBCT图像质量进行两次检查。使用磁共振成像(MR)和另一个体模的计算机断层扫描(CT)图像研究了使用CBCT图像进行图像配准的准确性。使用带有拟人模型的胶片剂量测定法在整个过程测试中测量整体位置精度。辐射等中心线在中心和极端位置的位置误差均小于0.1mm。 CBCT坐标系的三维偏差稳定了一年半(平均0.04±0.02 mm)。图像校正显示,CT和CBCT图像之间相差0.2±0.1 mm,而MR和CBCT图像之间相差0.4±0.2 mm。基于面罩的照射位置精度的整个过程测试显示精度为0.5±0.6mm。辐射等中心线精度,患者卧榻移动精度和Gamma Knife Icon CBCT精度均约为0.1毫米,并且可以稳定运行一年半。通过基调分配给MR图像的坐标系统比基准标记定义的系统更精确。在评估无框GKRS的整体准确性时,应考虑患者在照射过程中可能会运动。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号