Polymer gel dosimeter BANG3(R) was used to measure beam placement accuracy of a Leksell Gamma Knife radiosurgery unit. The authors did two experiments over a five-month period. Both show a 1.8-mm (corresponding to the size of two pixels) shift of measured dose distributions relative to the isodose lines calculated by the treatment planning software in the left-right direction. There is no observable error in the anterior-posterior direction. The authors have considered several sources for the observed shift, which include phantom setup error, MRI image distortion due to field inhomogeneity and non-linearity, treatment planning software error, and design/manufacturing error of the radiosurgery unit. The observed error has not been identified using the quality assurance procedure recommended by the manufacturer. The authors have demonstrated, therefore, that the polymer gel dosimeter is useful to verify the beam placement accuracy of a Leksell Gamma Knife unit as a part of acceptance test of hardware and treatment planning software. They recommend that the treatment prescription should be made considering the accuracy limit presented in this study.
机译:聚合物凝胶剂量计BANG3用于测量Leksell Gamma刀放射外科设备的射束放置精度。作者在五个月的时间内进行了两个实验。两者均显示相对于由治疗计划软件在左右方向上计算出的等剂量线,测量剂量分布有1.8毫米(对应于两个像素的大小)的偏移。在前后方向上没有可观察到的误差。作者考虑了观察到的偏移的几种来源,包括体模设置错误,由于场不均匀和非线性导致的MRI图像失真,治疗计划软件错误以及放射外科单元的设计/制造错误。使用制造商建议的质量保证程序尚未发现观察到的错误。因此,作者证明了聚合物凝胶剂量计可用于验证Leksell伽玛刀单元的射束放置精度,作为硬件和治疗计划软件验收测试的一部分。他们建议应考虑本研究中提出的准确度限值来制定治疗处方。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号