Remote dosimetric auditing of clinical trials: The need for vendor specific models to convert images to dose

机译:临床试验的远程剂量学审核:需要特定于供应商的模型以将图像转换为剂量

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

IntroductionA previous pilot study has demonstrated the feasibility of a novel image‐based approach for remote dosimetric auditing of clinical trials. The approach uses a model to convert in‐air acquired intensity modulated radiotherapy (IMRT) images to delivered dose inside a virtual phantom. The model was developed using images from an electronic portal imaging device (EPID) on a Varian linear accelerator. It was tuned using beam profiles and field size factors (FSFs) of a series of square fields measured in water tank. This work investigates the need for vendor specific conversion models for image‐based auditing. The EPID measured profile and FSF data for Varian (vendor 1) and Elekta (vendor 2) systems are compared along with the performance of the existing Varian model (VM) and a new Elekta model (EM) for a series of audit IMRT fields measured on vendor 2 systems.
机译:简介先前的一项先导研究已经证明了一种基于图像的新颖方法用于临床试验的远程剂量测定审计的可行性。该方法使用模型将空中获取的强度调制放射治疗(IMRT)图像转换为虚拟体模内的给药剂量。该模型是使用Varian线性加速器上的电子门禁成像设备(EPID)的图像开发的。它是使用光束轮廓和在水箱中测量的一系列正方形场的场大小因子(FSF)进行调整的。这项工作调查了针对基于映像的审核对特定于供应商的转换模型的需求。比较了针对Varian(供应商1)和Elekta(供应商2)系统的EPID测量的配置文件和FSF数据,以及针对一系列审计IMRT字段的现有Varian模型(VM)和新Elekta模型(EM)的性能在供应商2系统上。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号