Central Blood Pressure Monitoring via a Standard Automatic Arm Cuff

机译:通过标准的自动袖套监测中央血压

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

Current oscillometric devices for monitoring central blood pressure (BP) maintain the cuff pressure at a constant level to acquire a pulse volume plethysmography (PVP) waveform and calibrate it to brachial BP levels estimated with population average methods. A physiologic method was developed to further advance central BP measurement. A patient-specific method was applied to estimate brachial BP levels from a cuff pressure waveform obtained during conventional deflation via a nonlinear arterial compliance model. A physiologically-inspired method was then employed to extract the PVP waveform from the same waveform via ensemble averaging and calibrate it to the brachial BP levels. A method based on a wave reflection model was thereafter employed to define a variable transfer function, which was applied to the calibrated waveform to derive central BP. This method was evaluated against invasive central BP measurements from patients. The method yielded central systolic, diastolic, and pulse pressure bias and precision errors of −0.6 to 2.6 and 6.8 to 9.0 mmHg. The conventional oscillometric method produced similar bias errors but precision errors of 8.2 to 12.5 mmHg (p ≤ 0.01). The new method can derive central BP more reliably than some current non-invasive devices and in the same way as traditional cuff BP.
机译:当前的用于监测中心血压(BP)的示波设备将袖带压力保持在恒定水平,以获取脉搏体积体积描记(PVP)波形并将其校准为通过人群平均方法估算的肱动脉BP水平。开发了一种生理学方法来进一步推进中心血压测量。应用一种特定于患者的方法,通过常规的通气过程中通过非线性动脉顺应性模型获得的袖带压力波形来估算肱动脉血压水平。然后采用生理学启发的方法,通过整体平均从同一波形中提取PVP波形,并将其校准至肱动脉BP水平。此后,采用基于波反射模型的方法来定义变量传递函数,将其应用于校准波形以得出中心BP。针对来自患者的侵入性中央血压测量评估了该方法。该方法产生的中心收缩压,舒张压和脉搏压力偏差以及-0.6至2.6mmHg和6.8至9.0mmHg的精确度误差。传统的示波法产生类似的偏置误差,但精度误差为8.2至12.5mmHg(p≤0.01)。与传统的无创器械相比,该新方法可比某些当前的非侵入性设备更可靠地获得中心血压。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号