Quantitative detection of nitric oxide in exhaled human breath by extractive electrospray ionization mass spectrometry

机译:萃取电喷雾电离质谱法定量检测呼出气中一氧化氮

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

Exhaled nitric oxide (eNO) is a useful biomarker of various physiological conditions, including asthma and other pulmonary diseases. Herein a fast and sensitive analytical method has been developed for the quantitative detection of eNO based on extractive electrospray ionization mass spectrometry (EESI-MS). Exhaled NO molecules selectively reacted with 2-phenyl-4, 4, 5, 5-tetramethylimidazoline-1-oxyl-3-oxide (PTIO) reagent, and eNO concentration was derived based on the EESI-MS response of 1-oxyl-2-phenyl-4, 4, 5, 5-tetramethylimidazoline (PTI) product. The method allowed quantification of eNO below ppb level (~0.02 ppbv) with a relative standard deviation (RSD) of 11.6%. In addition, eNO levels of 20 volunteers were monitored by EESI-MS over the time period of 10 hrs. Long-term eNO response to smoking a cigarette was recorded, and the observed time-dependent profile was discussed. This work extends the application of EESI-MS to small molecules (<30 Da) with low proton affinity and collision-induced dissociation efficiency, which are usually poorly visible by conventional ion trap mass spectrometers. Long-term quantitative profiling of eNO by EESI-MS opens new possibilities for the research of human metabolism and clinical diagnosis.
机译:呼出气一氧化氮(eNO)是各种生理状况(包括哮喘和其他肺部疾病)的有用生物标记。在此,已经开发了一种快速灵敏的分析方法,用于基于萃取电喷雾电离质谱(EESI-MS)的eNO定量检测。呼出的NO分子与2-苯基-4、4、5、5-四甲基咪唑啉-1-氧基1-3氧化物(PTIO)选择性反应,根据1-氧基1-2的EESI-MS响应推导出eNO浓度-苯基-4、4、5、5-四甲基咪唑啉(PTI)产物。该方法可以定量分析低于ppb水平(〜0.02〜ppbv)的eNO,相对标准偏差(RSD)为11.6%。此外,在10分钟内,通过EESI-MS监测了20名志愿者的eNO水平。记录了对吸烟的长期eNO反应,并讨论了观察到的时间依赖性。这项工作将EESI-MS的应用扩展到了具有低质子亲和力和碰撞诱导解离效率的小分子(<30 Da),这通常在常规离子阱质谱仪中很难看到。 EESI-MS对eNO进行长期定量分析为人类代谢和临床诊断的研究提供了新的可能性。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号