Effect of electrostatic charge flow delay and multiple actuations on the in vitro delivery of salbutamol from different small volume spacers for infants.

机译:静电荷流量延迟和多次驱动对沙丁胺醇从婴幼儿不同小容量垫片体外递送的影响。

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

BACKGROUND: A study was undertaken to determine the influences of electrostatic charge, flow, delay, and multiple actuations on the in vitro delivery of salbutamol generated by a pressurised metered dose inhaler (pMDI) from small volume spacers used in infants. METHODS: Ten actuations from a salbutamol pMDI were drawn at different flow rates after either single or multiple actuations, with or without delay, through either static or reduced static spacers. An ionic detergent was used to reduce the charge of plastic spacers (Babyhaler, Babyspacer, Aerochamber, Nebuhaler). Electrostatic charge was measured using an electrometer. A multistage liquid impinger was used to determine the particle size distribution of the output of the pMDI through the spacers. RESULTS: Electrostatic charge on the surface of plastic spacers had the greatest influence on delivery, causing a decrease in drug delivery. Reducing charge by coating the surface with ionic detergent resulted in an increase of 46.5-71.1% (p < 0.001) in small (< 6.8 microns) particle delivery from small volume plastic spacers. Lower flow, delay, and multiple actuations resulted in decreased delivery from static spacers. Lower flow resulted in a decrease of 15% in small (< 9.6 microns) particle delivery. Delay and multiple actuations resulted in a decrease of 40.7% and 76.0%, respectively, in small (< 6.8 microns) particle delivery. The influences of lower flow, delay, and multiple actuations were greatly reduced or even eliminated by reducing charge. However, multiple actuations still resulted in a significant decreased delivery (p < 0.05). The reduced static Nebuhaler had a higher delivery than all small volume spacers. CONCLUSIONS: Electrostatic charge has a major influence on the delivery of salbutamol from small volume spacers. Using a metal spacer or ionic detergent coating of plastic spacers resulted in no or reduced charge and hence in improved delivery. Lower flow, delay, and multiple actuations played a major part only in static spacers.
机译:背景:进行了一项研究以确定静电荷,流量,延迟和多次启动对婴儿中使用的小剂量间隔物加压计量吸入器(pMDI)产生的沙丁胺醇的体外递送的影响。方法:单次或多次驱动后,通过沙丁胺醇pMDI的十次驱动以不同的流速通过静态或减少静态垫片进行延迟或不延迟的驱动。离子型清洁剂用于减少塑料垫片(Babyhaler,Babyspacer,Aerochamber,Nebuhaler)的电荷。使用静电计测量静电电荷。使用多级液体冲击器来确定通过间隔物的pMDI输出的粒度分布。结果:塑料垫片表面上的静电对传递的影响最大,从而导致药物传递的减少。通过用离子去污剂涂覆表面来减少电荷,这使得从小体积塑料垫片中获得的小颗粒(<6.8微米)增加了46.5-71.1%(p <0.001)。较低的流量,延迟和多次驱动导致静态垫片的输送减少。较低的流量导致小(<9.6微米)颗粒输送减少了15%。延迟和多次驱动分别导致小(<6.8微米)颗粒输送减少了40.7%和76.0%。降低流量,延迟和多次驱动的影响已通过减少电荷而大大减少甚至消除了。但是,多次驱动仍然导致输送量显着下降(p <0.05)。减少了静电的Nebuhaler的输送量高于所有小体积的垫片。结论:静电电荷对小剂量间隔物中沙丁胺醇的递送有重要影响。使用金属间隔物或塑料间隔物的离子去污剂涂层不会产生电荷,也不会减少电荷,因此改善了输送效果。较低的流量,延迟和多次驱动仅在静态垫片中起主要作用。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号