Noise power properties of a cone-beam CT system for breast cancer detection

机译:用于乳腺癌检测的锥束CT系统的噪声功率特性

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

The noise power properties of a cone-beam computed tomography (CT) system dedicated for breast cancer detection were investigated. Uniform polyethylene cylinders of various diameters were scanned under different system acquisition conditions. Noise power spectra were calculated from difference data generated by subtraction between two identical scans. Multidimensional noise power spectra (NPS) were used as the metric to evaluate the noise properties of the breast CT (bCT) under different system acquisition and reconstruction conditions. A comprehensive investigation of the noise properties was performed in regard to system acquisition parameters including kVp, mA, number of cone-beam projection images used, cone angle, and object size. The influence on reconstruction parameters including interpolation method, reconstruction filter, field of view, matrix size, and slice thickness were also studied. Under certain conditions, the zero-dimensional NPS (image variance) was used as a quantitative index to compare the influence from different scan parameters, especially the radiation dose. If the total scan dose is changed by linearly changing the total number of projection images while the dose per frame is kept constant, the noise power has a linear relationship with the reciprocal of the total dose. If the total scan dose is changed by linearly changing the dose per frame while the total number of projection images is kept constant, the noise power has a quadratic relationship with the reciprocal of the total dose. With the same amount of total dose, using fewer projection images results in lower image noise power in the CT image. Quantitative results from this noise power analysis provide guidance for the bCT system operation, optimization, and data reconstruction.
机译:研究了专用于乳腺癌检测的锥束计算机断层扫描(CT)系统的噪声功率特性。在不同的系统采集条件下扫描各种直径的均匀聚乙烯圆柱体。噪声功率谱是根据两次相同扫描之间相减得出的差异数据计算得出的。多维噪声功率谱(NPS)被用作评估在不同系统采集和重建条件下乳房CT(bCT)的噪声特性的指标。关于系统采集参数,包括kVp,mA,使用的锥束投影图像数,锥角和物体尺寸,对噪声特性进行了全面研究。还研究了对重构参数的影响,包括插值方法,重构滤波器,视场,矩阵大​​小和切片厚度。在某些条件下,将零维NPS(图像方差)用作定量指标,以比较不同扫描参数(尤其是辐射剂量)的影响。如果在每帧剂量不变的情况下通过线性改变投影图像的总数来改变总扫描剂量,则噪声功率与总剂量的倒数具有线性关系。如果在使投影图像的总数保持恒定的同时通过线性改变每帧的剂量来改变总扫描剂量,则噪声功率与总剂量的倒数具有二次关系。在总剂量相同的情况下,使用较少的投影图像会导致CT图像中的图像噪声功率降低。噪声功率分析的定量结果为bCT系统的操作,优化和数据重建提供了指导。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号