首页> 美国卫生研究院文献>Medical Physics >The effect of scatter and glare on image quality in contrast-enhanced breast imaging using an a-Si∕CsI(Tl) full-field flat panel detector
【2h】

The effect of scatter and glare on image quality in contrast-enhanced breast imaging using an a-Si∕CsI(Tl) full-field flat panel detector

机译:使用a-Si ∕ CsI(Tl)全场平板探测器在增强对比的乳腺成像中散射和眩光对图像质量的影响

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

The purpose of this study is to evaluate the performance of an antiscatter grid and its potential benefit on image quality for a full-field digital mammography (FFDM) detector geometry at energies typical for temporal subtraction contrast-enhanced (CE) breast imaging. The signal intensities from primary, scatter, and glare were quantified in images acquired with an a-Si∕CsI(Tl) FFDM detector using a Rh target and a 0.27 mm Cu filter at tube voltages ranging from 35 to 49 kV. Measurements were obtained at the center of the irradiation region of 20–80 mm thick breast-equivalent phantoms. The phantoms were imaged with and without an antiscatter grid. Based on these data, the performance of the antiscatter grid was determined by calculating the primary and scatter transmission factors (TP and TS) and Bucky factors (Bf). In addition, glare-to-primary ratios (GPRs) and scatter-to-primary ratios (SPRs) were quantified. The effect of the antiscatter grid on the signal-difference-to-noise ratio (SDNR) was also assessed. It was found that TP increases with kV but does not depend on the phantom thickness; TP values between 0.81 and 0.84 were measured. TS increases with kV and phantom thickness; TS values between 0.13 and 0.21 were measured. Bf decreases with kV and increases with phantom thickness; Bf ranges from 1.4 to 2.1. GPR is nearly constant, varying from 0.10 to 0.11. SPR without an antiscatter grid (SPR) ranges from 0.35 to 1.34. SPR decreases by approximately 9% from 35 to 49 kV for a given phantom thickness and is 3.5 times larger for an 80 mm thick breast-equivalent phantom than for a 20 mm thick breast-equivalent phantom. SPR with an antiscatter grid (SPR+) ranges from 0.06 to 0.31. SPR+ increases by approximately 23% from 35 to 49 kV for a given phantom thickness; SPR+ is four times larger for an 80 mm breast-equivalent phantom than for a 20 mm breast-equivalent phantom. When imaging a 25 mm PMMA plate at the same mean glandular dose with and without an antiscatter grid, the SDNR is 4% greater with a grid than without. For an 75mm PMMA plate, the SDNR is 20% greater with a grid. In conclusion, at the higher x-ray energy range used for CE-DM and CE-DBT, an antiscatter grid significantly reduces SPR and improves SDNR. These effects are most pronounced for thick breasts.
机译:这项研究的目的是评估反散射栅格的性能及其对全场数字乳房X线摄影(FFDM)检测器几何形状在图像质量(时间减影对比增强(CE)乳房成像的典型能量)下的潜在优势。在35至49 kV的管电压下,使用Rh目标和0.27 mm铜滤波器,使用a-Si ∕ CsI(Tl)FFDM检测器采集的图像中定量了来自初级,散射和眩光的信号强度。在20-80毫米厚的相当于乳房的幻影的照射区域中心进行测量。体模在有和没有反散射网格的情况下成像。根据这些数据,通过计算主要和散射传输因子(TP和TS)和Bucky因子(Bf)来确定防散射网格的性能。此外,还对眩光与主要比例(GPR)和散射与主要比例(SPR)进行了量化。还评估了防散射网格对信噪比(SDNR)的影响。发现TP随kV增加而增加,但不取决于体模厚度。测得的TP值在0.81和0.84之间。 TS随着kV和幻像厚度的增加而增加;测量的 T S 值介于0.13和0.21之间。 B f 随kV减小,随幻像厚度增加; B f 的范围是1.4到2.1。 GPR 几乎恒定,范围从0.10到0.11。没有反散射网格( SPR -)的 SPR 范围为0.35至1.34。对于给定的幻像厚度, SPR -从35 kV降低至49 kV约降低9%,对于80毫米厚的相当于乳房的幻影,其20毫米的幻影是3.5倍大相当于乳房的厚幻影。具有反散射网格( SPR + )的 SPR 范围为0.06至0.31。对于给定的幻像厚度, SPR + 从35 kV增加到49 kV约增加23%; SPR + 对于等同于80毫米乳房的幻象来说,是20毫米等同于乳房的幻象的四倍。在有和没有防散射栅格的情况下,以相同的平均腺体剂量对25 mm PMMA板成像时, SDNR 栅格比无栅格时大4%。对于75mm PMMA板,带网格的 SDNR 大20%。总之,在用于CE-DM和CE-DBT的更高的X射线能量范围内,防散射栅大大降低了 SPR 并改善了 SDNR 。这些效果对于丰满的乳房最为明显。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号