Computer simulation methods were used to evaluate the spatial resolution properties of a dedicated cone-beam breast CT system. x-ray projection data of a 70 µm nickel-chromium wire were simulated. The modulation transfer function (MTF) was calculated from the reconstructed axial images at different radial positions from the isocenter to study the spatial dependency of the spatial resolution of the breast CT scanner. The MTF was also calculated in both the radial and azimuthal directions. Subcomponents of the cone beam CT system that affect the MTF were modeled in the computer simulation in a serial manner, including the x-ray focal spot distribution, gantry rotation under the condition of continuous fluoroscopy, detector lag, and detector spatial resolution. Comparison between the computer simulated and physically measured MTF values demonstrates reasonable accuracy in the simulation process, with a small systematic difference (~9.5±6.4% difference, due to unavoidable uncertainties from physical measurement and system calibration). The intrinsic resolution in the radial direction determined by simulation was about 2.0 mm−1 uniformly through the field of view. The intrinsic resolution in the azimuthal direction degrades from 2.0 mm−1 at the isocenter to 1.0 mm−1 at the periphery with 76.9 mm from the isocenter. The results elucidate the intrinsic spatial resolution properties of the prototype breast CT system, and suggest ways in which spatial resolution can be improved with system modification.
机译:使用计算机仿真方法来评估专用锥束乳房CT系统的空间分辨率特性。模拟了一条70 µm镍铬丝的X射线投影数据。从等中心点不同径向位置处的重建轴向图像中计算出调制传递函数(MTF),以研究乳房CT扫描仪空间分辨率的空间依赖性。还沿径向和方位角方向计算了MTF。在计算机模拟中以串行方式对影响MTF的锥束CT系统的子组件进行建模,包括X射线焦点分布,连续荧光检查条件下的机架旋转,探测器滞后和探测器空间分辨率。计算机仿真的MTF值和物理测量的MTF值之间的比较表明,在仿真过程中具有合理的准确性,并且系统差异很小(由于物理测量和系统校准不可避免地存在不确定性,差异约为9.5±6.4%)。通过模拟确定的径向固有分辨率在整个视场中均匀约为2.0 mm -1 。方位方向上的本征分辨率从等角点处的2.0 mm -1 降低到等角点处76.9 mm的外围的1.0 mm -1 。结果阐明了原型乳腺CT系统的固有空间分辨率特性,并提出了通过系统修改可以改善空间分辨率的方法。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号