Sensorless Force Sensing for Minimally Invasive Surgery

机译:微创手术的无传感器力感测

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

Robotic minimally invasive surgery (R-MIS) has achieved success in various procedures; however, the lack of haptic feedback is considered by some to be a limiting factor. The typical method to acquire tool–tissue reaction forces is attaching force sensors on surgical tools, but this complicates sterilization and makes the tool bulky. This paper explores the feasibility of using motor current to estimate tool-tissue forces and demonstrates acceptable results in terms of time delay and accuracy. This sensorless force estimation method sheds new light on the possibility of equipping existing robotic surgical systems with haptic interfaces that require no sensors and are compatible with existing sterilization methods.
机译:机器人微创手术(R-MIS)在各种手术中均取得了成功。然而,一些人认为缺乏触觉反馈是一个限制因素。获取工具-组织反作用力的典型方法是在手术工具上安装力传感器,但这会使灭菌复杂化,并使工具笨重。本文探讨了使用电动机电流来估计工具组织力的可行性,并就时间延迟和准确性证明了可接受的结果。这种无传感器的力估计方法为现有的机器人手术系统配备了不需要传感器且与现有灭菌方法兼容的触觉界面的可能性提供了新的思路。