In Vitro Performance Assessment of Distal Protection Filters: Pulsatile Flow Conditions

机译:远端保护过滤器的体外性能评估:脉动流量条件

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

>Purpose: To evaluate in vitro the capture efficiency of a distal protection filter (DPF) used during carotid artery stenting and examine the longitudinal vascular impedance in the presence of a DPF.>Methods: Four approved DPFs (Spider RX, FilterWire EZ, RX Accunet, and FiberNet) were evaluated in a physiologically realistic in vitro setup. A pulsatile programmable piston pump circulated a blood analog at a time-varying flow rate representative of the human common carotid artery. A silicone carotid bifurcation having average human dimensions was used for the carotid flow model. Microspheres ranging from 40 to 900 µm were injected to simulate embolization. The longitudinal vascular impedance was calculated as the ratio of the time-varying pressure gradient across the DPF to the time-varying flow rate in the internal carotid artery.>Results: RX Accunet had the highest capture efficiency (99.4%) and Spider RX the lowest (78.1%). Spider RX increased the longitudinal vascular impedance the least after deployment (+23%), while FilterWire EZ increased the longitudinal vascular impedance the least after particles were injected (+29%). FiberNet increased longitudinal vascular impedance the most (+84%).>Conclusion: This investigation, unique for examining the effects of physiologically realistic pulsatile flow on DPF performance, can aid in the development of future generations of novel DPFs.
机译:>目的:要在体外评估在颈动脉支架置入术中使用的远端保护滤器(DPF)的捕获效率,并在存在DPF的情况下检查纵向血管阻抗。>方法:在生理上可行的体外设置中评估了四个批准的DPF(Spider RX,FilterWire EZ,RX Accunet和FiberNet)。脉动可编程活塞泵以代表人类颈总动脉的时变流速循环血液类似物。将具有人类平均尺寸的硅酮颈动脉分叉用于颈动脉血流模型。注入范围为40至900 µm的微球以模拟栓塞。纵向血管阻抗的计算方法是:DPF上的时变压力梯度与颈内动脉中时变流速的比率。>结果: RX Accunet的捕获效率最高(99.4) %),而Spider RX最低(78.1%)。 Spider RX在展开后至少增加了纵向血管阻抗(+ 23%),而FilterWire EZ在注射粒子之后增加了纵向血管阻抗(+ 29%)。 FiberNet增加了最大的纵向血管阻抗(+84%)。>结论:该研究是检查生理上真实的脉动血流对DPF性能的影响的独特方法,可以帮助开发新一代DPF。 。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号