Light Intensity Decay in Quartz-Tungsten-Halogen Polymerization Units

机译:石英-钨-卤素聚合装置中的光强衰减

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

>Objectives: The aim of this study was to determine the effect of ageing on the light intensity of quartz-tungsten-halogen (QTH) polymerization units in dental schools and dental clinics.>Materials & Methods: Two dental schools and multiple public health dental clinics were visited and the light intensity of curing units was assessed. The age, make, model and service history were recorded. A calibrated radiometer (range 0-1000 nm) was used to measure the light intensity of 171 QTH units. At the end of each visit, the team members provided each dental clinic with a verbal report of the test results and recommendations on maintenance of light polymerization units.>Results: The 171 units tested were from seven different manufacturers. The mean light intensity was 373.1 mW/cm2, the lowest intensity being 95.3 and the highest 526.60 mW/cm2. Only 29 QTH units recorded more than 500 mW/cm2; 102 units showed readings that ranged between 300 and 500 mW/cm2 and 39 units recorded readings below 300 mW/cm2. It was not able that one unit had light intensity reading of less than 100 mW/cm2. The age distribution was as follows: 35 units were more than 10 years old, 79 were from 5 to 10 years old and 57 were less than 5 years old. There was a high statistical significant relation between the mean light intensity and age (P = 0.001).>Conclusion: Light intensity values used in some dental clinics were to be found lower than what is recommended for optimum polymerization of resin-based composite restorations. Awareness among dentists of the value of QTH light curing units and their commitment to care and maintenance are crucial to achieving optimum restoration properties.>How to cite this article: Al-Samadani KH, Al-Dharrab A, Wahbi MA , Algizani L. Light Intensity Decay in Quartz-Tungsten-Halogen Polymerization Units. J Int Oral Health 2013; 5(1):23-30.
机译:>目标:本研究的目的是确定老化对牙科学校和牙科诊所中石英-钨-卤素(QTH)聚合单元的光强度的影响。>材料与方法: 访问了两家牙科学校和多家公共卫生牙科诊所,并评估了固化单元的光强度。记录了年龄,品牌,型号和服务历史。使用校准的辐射计(范围为0-1000 nm)来测量171个QTH单位的光强度。在每次拜访结束时,团队成员向每个牙科诊所提供有关测试结果的口头报告,以及有关维护轻聚合单元的建议。>结果:测试的171个单元来自七个不同的制造商。平均光强为373.1 mW / cm 2 ,最低为95.3,最高为526.60 mW / cm 2 。只有29个QTH单位记录的功率大于500 mW / cm 2 ; 102个单位的读数介于300至500 mW / cm 2 之间,而39个单位的读数低于300 mW / cm 2 。一个单元的光强度读数不能低于100 mW / cm 2 。年龄分布如下:超过10年的有35个单位,从5至10年的有79个单位,小于5岁的有57个单位。平均光强度与年龄之间存在极高的统计学显着关系(P = 0.001)。>结论:发现某些牙科诊所使用的光强度值低于最佳聚合强度的建议值。树脂基复合修复体。牙医对QTH光固化机的价值及其对维护和保养的承诺的认识对于实现最佳修复性能至关重要。>如何引用本文:Al-Samadani KH,Al-Dharrab A,Wahbi MA,Algizani L.石英-钨-卤素聚合单元中的光强度衰减。 《国际口腔卫生杂志》 2013; 5(1):23-30。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号