A rebreathing method for measuring carbon monoxide transfer factor (TLCO) and transfer coefficient (KCO) is evaluated. The test was performed on 42 subjects. Seventeen of these had normal lung function while the remainder suffered from various conditions, most having fibrosing alveolitis. No patients with severe airways obstruction were included in the group. The results were compared with single-breath measurements and close correlations were found between single-breath KCO and rebreathing KCO (r = 0.78, p less than 0.001) and between single-breath TLCO and rebreathing TLCO (r = 0.97, p less than 0.001). The reasons for discrepancies between single-breath and rebreathing values are discussed. It is concluded that measurement of rebreathing gas transfer is a useful supplement to routine tests of lung function, particularly in breathless patients with a small vital capacity.
机译:评价了用于测量一氧化碳传递因子(TLCO)和传递系数(KCO)的呼吸方法。该测试对42位受试者进行。其中的十七个肺功能正常,其余的则患有各种疾病,大多数患有纤维化性肺泡炎。该组中没有严重气道阻塞的患者。将结果与单呼吸测量进行比较,发现单呼吸KCO和再呼吸KCO之间的相关性(r = 0.78,p小于0.001)和单呼吸TLCO和再呼吸TLCO之间的相关性(r = 0.97,p小于0.001) )。讨论了单次呼吸值和重新呼吸值之间差异的原因。结论是呼吸气体转移的测量是常规肺功能检查的有用补充,特别是对于肺活量小的呼吸困难患者。