摘要

The ability of 12 commercially available cardiotomy reservoirs to remove bubbles from aspirated blood was investigated by means of a simulated cardiopulmonary bypass circuit and an ultrasonic microbubble detector. Performance varied considerably. The number of gaseous microemboli remaining after passage of blood through the reservoir was reduced by (a) holding the blood in the reservoir, (b) reducing the volume of air mixed with the aspirated blood, and (c) using a reservoir that did not induce turbulence and that contained integral micropore filtration material. Further micropore filtration of the blood after passage through the cardiotomy reservoir was beneficial, and significantly more bubbles were extracted when the microfilter was sited below the reservoir than when it was placed in the arterial line.
机译:通过模拟的心肺旁路电路和超声微气泡检测器,研究了12种市售的心脏切开术储存器从吸入的血液中去除气泡的能力。性能差异很大。通过(a)将血液保存在容器中,(b)减少与吸入的血液混合的空气量,以及(c)使用没有通过容器的血液来减少血液通过容器后残留的气态微栓子的数量。引起湍流,并且包含整体微孔过滤材料。通过心脏切开术储液器后,进一步对血液进行微孔过滤是有益的,当将微滤器置于储液器下方时,与放置在动脉管线中相比,提取出的气泡明显更多。