The purpose of our study was to clarify limitations of off-label use for low cost nonmedical use (NMU) pulse oximeters by primary care providers. These devices are widely marketed over the Internet and in drugstores but are not intended for medical use or reviewed by the Food and Drug Administration (FDA). Our study compared oxygen saturation (SpO2) in patients from 1 medical use (MU) pulse oximeter to 8 NMU pulse oximeters. Measured arterial oxygenation (SaO2) was compared with SpO2 when available. In patients who were normoxic (SpO2 ≥90%), all oximeters exhibited similar readings. This finding suggests that NMU pulse oximeters may be able to rule out hypoxemia in clinical settings.
机译:我们研究的目的是阐明初级保健提供者对廉价非医疗用途(NMU)脉搏血氧仪进行标签外使用的局限性。这些设备已在Internet和药店中广泛销售,但不适用于医疗用途或未经美国食品药品监督管理局(FDA)审查。我们的研究比较了从1台医用(MU)脉搏血氧仪到8台NMU脉搏血氧仪的患者的血氧饱和度(SpO2)。如果可用,将测得的动脉氧合(SaO2)与SpO2进行比较。在常氧(SpO2≥90%)患者中,所有血氧仪均显示相似的读数。这一发现表明,NMU脉搏血氧仪可以在临床环境中排除低氧血症。