Evaluation of the accuracy of non-invasive automatic blood pressure monitors.

机译:评估无创自动血压计的准确性。

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

Non-invasive automatic blood pressure monitors (BP-103N, DINAMAP 845XT, Finapres 2300) were compared with the auscultatory method. The blood pressure readings given by the oscillometric method (BP-103N, DINAMAP 845XT) were accurate and reproducible. Agreement with the auscultatory method was especially good for systolic pressure. For diastolic pressure readings, there was less agreement with the results of the auscultatory method. The finger arterial pressure method (Finapres 2300) occasionally displayed greater variability than the devices using the oscillometric method.
机译:将无创自动血压监测仪(BP-103N,DINAMAP 845XT,Finapres 2300)与听诊法进行了比较。示波法(BP-103N,DINAMAP 845XT)给出的血压读数是准确且可重复的。与听诊法相吻合尤其适合收缩压。对于舒张压读数,与听诊法的结果不一致。手指动脉压方法(Finapres 2300)偶尔会比使用示波法的设备显示更大的可变性。