Performance of low-pressure cuffs. An experimental evaluation.

机译:低压袖带的性能。实验评估。

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

A study was carried out in vitro to determine the minimum intracuff pressure needed to produce no-leak ventilation and prevent aspiration in Portex Profile and Searle Medical Sensiv endotracheal tubes. The mean pressures required to produced no-leak ventilation were 4.16 and 1.06 kPa respectively and the mean pressures at which aspiration occurred were 2.93 and 0.54 kPa respectively. The implications of these findings are discussed and recommendations are made regarding cuff specifications.
机译:进行了一项体外研究,以确定在Portex Profile和Searle Medical Sensiv气管导管中产生无泄漏通气并防止抽吸所需的最小袖带内压力。产生无泄漏通风所需的平均压力分别为4.16和1.06 kPa,发生抽吸的平均压力分别为2.93和0.54 kPa。讨论了这些发现的含义,并就袖带规格提出了建议。