Purpose: Hemodynamic performance of aortic valve bioprostheses is essential for reliable function and durability. So far, the supra-annularly implanted stentless Sorin Freedom Solo (SFS) demonstrated unsurpassed hemodynamic properties. As contemporary stented and externally mounted pericardial bioprostheses, like the Labcor Dokimos Plus (LDP), also improve hemodynamic performance, these types of valves were compared in this study.Methods: A total of 218 patients, who underwent aortic valve replacement with the LDP or the SFS, were matched retrospectively 1:1 on variables affecting hemodynamic measurements: implanted valve size, age, sex, and body surface area (BSA). With matching tolerance for valve size and gender of 0%, for age and BSA of 5%, 57 patient-pairs were yielded. Operative data, clinical, and hemodynamic outcome were analyzed.Results: Except for slightly higher left ventricular function and lower procedural times in the SFS group, preoperative, operative, and postoperative characteristics of patient-pairs did not differ significantly. Mean pressure gradients, effective orifice areas (EOAs), and indexed EOAs were comparable. Corresponding to valve sizes of 21, 23, 25, and 27 mm, the indexed EOAs of the LDP and SFS prostheses were 1.08 ± 0.33, 0.92 ± 0.19, 0.93 ± 0.24, 0.99 ± 0.13 cm2/m2 and 0.81 ± 0.13, 0.92 ± 0.28, 0.95 ± 0.20, 1.04 ± 0.27 cm2/m2, respectively.Conclusion: Contemporary stented and stentless pericardial bioprostheses showed excellent hemodynamic properties without significant differences in EOAs and indexed EOAs.
机译:目的:主动脉瓣生物假体的血流动力学性能对于可靠的功能和耐用性至关重要。到目前为止,超环植入的无支架Sorin Freedom Solo(SFS)表现出无与伦比的血液动力学特性。由于当代有支架和外部安装的心包生物假体(如Labcor Dokimos Plus(LDP))也可以改善血液动力学性能,因此在本研究中对这些类型的瓣膜进行了比较。方法:共有218例患者接受了LDP或LDP置换主动脉瓣在影响血流动力学测量的变量上,SFS进行了1:1匹配:植入的瓣膜大小,年龄,性别和体表面积(BSA)。瓣膜大小和性别的匹配公差为0%,年龄和BSA的匹配公差为5%,因此产生了57对患者。结果:除SFS组左心室功能略高和手术时间缩短外,患者对的术前,术中和术后特征无明显差异。平均压力梯度,有效孔口面积(EOA)和标定的EOA相当。对应于21、23、25和27 mm的瓣膜大小,LDP和SFS假体的分度EOA为1.08±0.33、0.92±0.19、0.93±0.24、0.99±0.13 cm 2 / m 2 和0.81±0.13,0.92±0.28,0.95±0.20,1.04±0.27 cm 2 / m 2 。支架式和无支架式心包生物假体显示出优异的血液动力学特性,而EOAs和分度的EOAs没有明显差异。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号