Effects of temperature and relative humidity on biological indicators used for ethylene oxide sterilization.

机译:温度和相对湿度对环氧乙烷灭菌所用生物指标的影响。

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

A study was made to determine the effects of temperature and moisture on the D-value of a common biological indicator. Relative humidity (RH) was varied between 10 and 70% in increments of 10%, and temperature was varied between 30 and 70 degrees C in increments of 10 degrees C. Temperature was found to have a pronounced effect on the D-value. At 60% RH, the D-value varied from 15.0 min at 30 degrees C to 1.1 min at 70 degrees C. When RH was plotted against the average D-value at the various temperatures, the temperature curves at or above 50 degrees C were more erratic and the RH had a significant effect. The study showed that temperature and RH must be controlled if biological indicators are to be properly calibrated for use in ethylene oxide sterilization.
机译:进行了一项研究以确定温度和湿度对常见生物指示剂D值的影响。相对湿度(RH)在10到70%之间以10%的增量变化,温度在30到70摄氏度之间以10℃的增量变化。发现温度对D值有显着影响。在60%RH下,D值从30摄氏度下的15.0分钟变化到70摄氏度下的1.1分钟。当相对于各种温度下的平均D值绘制RH时,在50摄氏度或以上的温度曲线为更加不稳定,相对湿度会产生重大影响。研究表明,要正确校准用于环氧乙烷灭菌的生物学指标,必须控制温度和相对湿度。