首页> 美国卫生研究院文献>Applied and Environmental Microbiology >Limulus amoebocyte lysate and direct sampling methods for surveillance of operating nebulizers.
【2h】

Limulus amoebocyte lysate and direct sampling methods for surveillance of operating nebulizers.

机译:a变形细胞溶解物和直接采样方法用于监视正在运行的雾化器。

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

The Limulus amoebocyte lysate test for detection of endotoxin (Pyrogent; Mallinckrodt Chemical Co.) and the Easicult method (Orion Diagnostica) for detection of bacteria were compared with direct dilution sampling, a standardized technique for respiratory therapy surveillance previously developed in our laboratory. Tests of 206 reservoirs of nebulizers were done in three hospitals in Georgia. Forty-five percent of all reservoirs sampled were contaminated. Gram-negative, nonfermentative bacilli were the predominant contaminants. The results of the Limulus test and the Easicult system were in agreement with those of the direct dilution sampling tests approximately 84 and 90% of the time, respectively. Direct dilution of water samples onto blood agar plates was the most sensitive, reliable, and informative method for detecting viable bacteria. The Easicult and Limulus systems were sensitive enough to detect greater than or equal to 10(3) colony-forming units per ml. Positive Limulus tests and negative culture tests, reflecting detection of endotoxin but not of viable gram-negative bacteria, occurred in 20 of 206 (9.7%) instances. Positive cultures and negative Limulus tests were noted in 13 of 206 (6.8%) samplings. The Limulus test is a valuable procedure, for it can detect moderate-to-heavy microbial contamination within 1 h of testing and affords the opportunity to remove contaminated equipment from patients within minutes of a positive test result. These results demonstrate the potential value of the Easicult and Limulus tests for selective surveillance of operating nebulizers.
机译:将用于检测内毒素的Li变形细胞溶解物检测(Pyrogent; Mallinckrodt Chemical Co.)和用于细菌检测的Easicult方法(Orion Diagnostica)与直接稀释采样进行了比较,直接稀释采样是我们实验室之前开发的用于呼吸治疗监控的标准化技术。在佐治亚州的三家医院对206个雾化器进行了测试。采样的所有水库中有百分之四十五受到污染。革兰阴性,非发酵杆菌是主要的污染物。 mul测试和Easicult系统的结果分别与直接稀释采样测试的结果一致,分别为84%和90%。将水样品直接稀释到血琼脂板上是检测活菌的最灵敏,最可靠和最有用的方法。 Easicult和Limulus系统足够灵敏,可以检测到每毫升大于或等于10(3)的菌落形成单位。在206例中有20例(9.7%)发生阳性Li试验和阴性培养试验,这反映了对内毒素的检测,但未检测到活的革兰氏阴性细菌。在206个样本中有13个(6.8%)记录了阳性培养和阴性negative测试。 Limulus测试是一种有价值的程序,因为它可以在测试的1小时内检测到中度至重度的微生物污染,并提供在阳性测​​试结果的几分钟内从患者身上移除受污染设备的机会。这些结果证明了Easicult和Limulus测试对于选择性监测正在运行的雾化器的潜在价值。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号