A comparison of transcutaneous bilirubinometers: SpectRx BiliCheck versus Minolta AirShields

机译:经皮胆红素计的比较:SpectRx BiliCheck与Minolta AirShields

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

>Background: Two devices are available for making transcutaneous estimates of serum bilirubin (SBR): the Minolta AirShields JM102 and the new SpectRx BiliCheck. >Objectives: (a) To measure how well the readings produced by these devices agree with SBR measured in the laboratory; (b) to estimate for each device, the proportion of infants with clinical jaundice who would require blood sampling if the device was used as a screening tool to detect infants with SBR ≥ 250 µmol/l. >Design: Prospective cohort study of jaundiced infants who required SBR at ≤ 20 days of postnatal age. Those who had received phototherapy or exchange transfusion were excluded. >Setting: Tertiary neonatal service in South-East Scotland. >Interventions: Within 30 minutes of SBR sampling, transcutaneous bilirubinometry was performed using one Minolta and two SpectRx devices (designated A and B). >Results: Sixty four neonates were enrolled, 19 of which were preterm (31–35 weeks). The 95% confidence intervals of a device reading corresponding to SBR were ± 66.7, ± 67.9, and ± 66.4 µmol/l respectively. Using the devices to identify all SBR ≥ 250 µmol/l would reduce SBR sampling by 23%, 16%, and 20% respectively. >Conclusions: Given that SBR levels range from 50 to 400 µmol/l in jaundiced infants, the 95% confidence intervals of the devices are wide at ± 67 µmol/l. The SpectRx can be used as a screening tool for hyperbilirubinaemia but there is no advantage in using it over the Minolta.
机译:>背景:有两种设备可用于透皮评估血清胆红素(SBR):Minolta AirShields JM102和新的SpectRx BiliCheck。 >目标:(a)测量这些设备产生的读数与实验室测量的SBR吻合程度; (b)估算每种设备,如果使用该设备作为筛查工具来检测SBR≥250μmol/ l的婴儿,需要抽血的临床黄疸婴儿的比例。 >设计:对出生后≤20天需要SBR的黄疸型婴儿进行前瞻性队列研究。那些接受光疗或换血的人被排除在外。 >设置:苏格兰东南部的第三级新生儿服务。 >干预措施:在SBR采样后的30分钟内,使用一台Minolta和两台SpectRx设备(称为A和B)进行了经皮胆红素测定。 >结果:招募了64例新生儿,其中19例早产儿(31-35周)。对应于SBR的设备读数的95%置信区间分别为±66.7,±67.9和±66.4 µmol / l。使用这些设备来识别所有≥250 µmol / l的SBR可以分别将SBR采样减少23%,16%和20%。 >结论:鉴于黄疸型婴儿的SBR水平范围为50至400 µmol / l,该装置的95%置信区间宽为±67 µmol / l。 SpectRx可用作高胆红素血症的筛查工具,但与Minolta相比,使用它没有优势。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号