Purpose. To evaluate Fourier-domain optical coherence tomography (FD-OCT) as an adjunct to traditional slit lamp examination of donor corneas with suspected Anterior Stromal Opacities. Methods. Seven corneas suspected of having anterior stromal opacities by slit lamp examination were evaluated with FD-OCT. Each cornea was evaluated to confirm the presence of opacity and, if present, the depth of opacity was measured. Results. The opacity depth ranged from 82 μm to 624 μm. The initial slit lamp impressions of five of the seven corneas were confirmed by OCT. In two corneas, the OCT findings were different from the initial slit lamp impressions. Slit lamp examination of the first cornea gave the impression of anterior stromal scarring, but OCT showed that the opacity was limited to the epithelium. Slit lamp examination of the second cornea suggested opacity limited to the epithelium, but OCT identified significant sub-Bowman's scarring. In all cases, the Eye Bank Technicians reported that the location and depth of corneal opacity were more sharply defined by OCT than by slit lamp. Conclusion. The high resolution of OCT makes it easier to determine the location of corneal opacities compared to slit lamp examinations. This enhanced visualization can improve decisions regarding transplant suitability of donor corneas.
机译:目的。评估傅立叶域光学相干断层扫描(FD-OCT)作为传统裂隙灯检查可疑前基质不透明的供体角膜的辅助手段。方法。用FD-OCT评估了通过裂隙灯检查怀疑有前间质混浊的七个角膜。评价每个角膜以确认不透明的存在,并且如果存在,则测量不透明的深度。结果。不透明深度范围为82μm至624μm。 OCT证实了七个角膜中五个角膜的初始裂隙灯压痕。在两个角膜中,OCT结果与最初的裂隙灯压印不同。第一个角膜的裂隙灯检查给人以前基质瘢痕的印象,但是OCT显示不透明仅限于上皮。裂隙灯检查第二个角膜表明不透明仅限于上皮,但OCT发现明显的Bowman瘢痕。在所有情况下,眼库技术人员报告说,OCT比裂隙灯更清楚地定义了角膜混浊的位置和深度。结论。与裂隙灯检查相比,OCT的高分辨率使确定角膜混浊的位置更加容易。这种增强的可视化功能可以改善有关供体角膜移植适合性的决策。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号