首页> 美国卫生研究院文献>BMC Musculoskeletal Disorders >Contact stresses pressure and area in a fixed-bearing total ankle replacement: a finite element analysis
【2h】

Contact stresses pressure and area in a fixed-bearing total ankle replacement: a finite element analysis

机译:固定支撑全踝关节置换中的接触应力压力和面积:有限元分析

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

BackgroundMobile-bearing ankle implants with good clinical results continued to increase the popularity of total ankle arthroplasty to address endstage ankle osteoarthritis preserving joint movement. Alternative solutions used fixed-bearing designs, which increase stability and reduce the risk of bearing dislocation, but with a theoretical increase of contact stresses leading to a higher polyethylene wear. The purpose of this study was to investigate the contact stresses, pressure and area in the polyethylene component of a new total ankle replacement with a fixed-bearing design, using 3D finite element analysis.
机译:背景具有良好临床效果的可移动的踝关节植入物继续增加全踝关节置换术的应用,以解决终末期踝关节骨关节炎,并保持关节运动。替代解决方案使用固定轴承设计,该设计可增加稳定性并降低轴承错位的风险,但理论上会增加接触应力,从而导致更高的聚乙烯磨损。这项研究的目的是使用3D有限元分析法研究固定支撑设计的新型全踝置换术中聚乙烯部件的接触应力,压力和面积。