首页> 美国卫生研究院文献>The British Journal of Ophthalmology >Minimally invasive direct real time measurement of drug concentration in the anterior eye
【2h】

Minimally invasive direct real time measurement of drug concentration in the anterior eye

机译:微创直接实时测量前眼中的药物浓度

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

>Aims: To evaluate a corneal contact lens which effectively turns the anterior chamber of the eye into a cuvette, enabling the concentration of a drug to be measured using absorption spectroscopy.>Methods: A hand held contact lens incorporating optical fibres connected to a spectrograph enabled a beam of light to be directed in, across, and out of the anterior chamber. The device was used to follow the time course of drug concentration in the anterior chamber of rabbit (sedated) and humans, using topical brimonidine or fluorescein (with or without local anaesthesia). Absorbance measurements were taken for a 5–25 second period, repeated every 30 minutes. Drug concentrations were compared using absorbance peak height.>Results: Corneal absorption starts to rise rapidly at wavelengths shorter than 315 nm. The light path within the anterior chamber is 6.9 mm (rabbit) and 5.8 mm (human), the absorbance measured also includes a corneal component. Application of fluorescein (three drops of 2% solution) in rabbit allowed detection, 60 minutes later, of a large absorbance peak at 490 nm. In the human eye, the device could not measure fluorescein (applied as in rabbit), but clearly detected brimonidine for 3 hours following topical application of 0.6 mg. Modification of the device to measure fluorescence resulted in the detection of 5.3 nM fluorescein in the ex vivo rabbit eye, an increase in sensitivity of two orders of magnitude over the absorption measurements.>Conclusion: This device has the potential to allow repeated measurements of drug concentrations in the anterior eye provided the drug has suitable absorption or fluorescence characteristics.
机译:>目的:要评估可有效将眼睛的前房变成比色皿的角膜接触镜,从而能够使用吸收光谱法测量药物的浓度。>方法:结合有与光谱仪连接的光纤的手持式隐形眼镜使光束能够被引导进入前房,穿过前房和从前房出来。使用局部溴莫尼定或荧光素(有或没有局部麻醉),使用该设备跟踪兔子(镇静剂)和人前房中药物浓度的时间变化。进行5–25秒的吸光度测量,每30分钟重复一次。使用吸光度峰高比较药物浓度。>结果:在短于315 nm的波长处,角膜吸收开始迅速增加。前房内的光路是6.9毫米(兔子)和5.8毫米(人),测得的吸光度还包括角膜成分。 60分钟后,在兔子中应用荧光素(3滴2%溶液)可以检测到490 nm处的大吸收峰。在人眼中,该设备无法测量荧光素(像在兔子中一样),但是在局部应用0.6 mg后3小时内可以清楚地检测到溴莫尼定。修改用于测量荧光的设备导致在离体兔眼中检测到5.3 nM荧光素,比吸收测量值的灵敏度提高了两个数量级。>结论:只要药物具有合适的吸收或荧光特性,就可以重复测量前眼中的药物浓度。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号