Clinical evaluation of scanning laser polarimetry: I Intraoperator reproducibility and design of a blood vessel removal algorithm

机译:扫描激光旋光仪的临床评估:术中可重复性和血管去除算法的设计

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

AIMS—To evaluate the reproducibility of the retardation values (change in polarisation) obtained with the scanning laser polarimeter in a series of normal subjects and glaucoma patients. To improve the analysis of the raw data by devising and evaluating a blood vessel removal algorithm.
METHODS—Scanning laser polarimetry was performed on 10 normal subjects and 10 glaucoma patients. A series of six images was obtained from each eye. The normal subjects were re-imaged 3 months after their initial assessment. The retardation values obtained from each eye were analysed using the authors' own methods, including the use of an algorithm to remove blood vessels from the polar profiles. The reproducibility of these measurements and the performance of the blood vessel removal algorithm were assessed.
RESULTS—The "individual point" coefficient of variation was approximately 12.5% for normal subjects and 17.0% for glaucoma patients. The "integral" coefficient of variation for these groups was approximately 5.5% and 9.5% respectively. The reproducibility of the measurements did not improve with an increased number of measurements. There was no difference in the reproducibility of the measurements in normal subjects over time. The blood vessel removal algorithm improved the reproducibility of the measurements when the shape of the profile was assessed.
CONCLUSION—The intraoperator reproducibility of retardation values obtained with the scanning laser polarimeter is satisfactory for its use as a clinical tool. The use of a blood vessel removal algorithm improves the reproducibility of the measurements and also assists the clinician in the interpretation of the polar profiles. Furthermore, it allows the construction of normal database polar profiles, thereby enabling the identification, location and quantification of retinal nerve fibre layer damage in an "at risk" individual's polar profile.

Keywords: scanning laser polarimetry; glaucoma; reproducibility; algorithm
机译:目的—为了评估在一系列正常受试者和青光眼患者中使用扫描激光偏振仪获得的延迟值(偏振变化)的可再现性。为了通过设计和评估血管去除算法来改善原始数据的分析。
方法-对10例正常受试者和10例青光眼患者进行了扫描激光极化仪。从每只眼睛获得一系列六个图像。正常受试者在初次评估后3个月重新成像。使用作者自己的方法分析从每只眼睛获得的延迟值,包括使用从极谱中去除血管的算法。评估了这些测量的可重复性和血管去除算法的性能。
结果-正常受试者的“个体点”变异系数约为12.5%,青光眼患者的变异系数约为17.0%。这些组的“积分”变异系数分别约为5.5%和9.5%。测量数量的增加并没有提高测量的可重复性。随着时间的推移,正常受试者的测量结果可重复性没有差异。当评估轮廓的形状时,血管去除算法提高了测量结果的可重复性。
结论—使用扫描激光旋光仪获得的手术中延迟值的术中可重复性对于用作临床工具是令人满意的。血管去除算法的使用提高了测量的可重复性,还帮助临床医生解释了极谱。此外,它还可以构建正常的数据库极化分布图,从而可以识别,定位和量化“处于危险中”个体的极化分布图中的视网膜神经纤维层损伤。

青光眼;重现性算法