Can CT scan protocols used for radiotherapy treatment planning be adjusted to optimize image quality and patient dose? A systematic review

机译:可以调整用于放射治疗计划的CT扫描方案以优化图像质量和患者剂量吗?系统评价

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

This article reviews publications related to the use of CT scans for radiotherapy treatment planning, specifically the impact of scan protocol changes on CT number and treatment planning dosimetry and on CT image quality. A search on PubMed and EMBASE and a subsequent review of references yielded 53 relevant articles. CT scan parameters significantly affect image quality. Some will also affect Hounsfield unit (HU) values, though this is not comprehensively reported on. Changes in tube kilovoltage and, on some scanners, field of view and reconstruction algorithms have been found to produce notable HU changes. The degree of HU change which can be tolerated without changing planning dose by >1% depends on the body region and size, planning algorithms, treatment beam energy and type of plan. A change in soft-tissue HU value has a greater impact than changes in HU for bone and air. The use of anthropomorphic phantoms is recommended when assessing HU changes. There is limited published work on CT scan protocol optimization in radiotherapy. Publications suggest that HU tolerances of ±20 HU for soft tissue and of ±50 HU for the lung and bone would restrict dose changes in the treatment plan to <1%. Literature related to the use of CT images in radiotherapy planning has been reviewed to establish the acceptable level of HU change and the impact on image quality of scan protocol adjustment. Conclusions have been presented and further work identified.
机译:本文回顾了与将CT扫描用于放射治疗治疗计划有关的出版物,特别是扫描协议更改对CT数量和治疗计划剂量学以及对CT图像质量的影响。在PubMed和EMBASE上进行搜索以及对参考文献的后续审查产生了53条相关文章。 CT扫描参数会严重影响图像质量。尽管这尚未得到全面报道,但也会影响Hounsfield单位(HU)的值。发现管千伏的变化以及在某些扫描仪上的视野和重建算法会产生显着的HU变化。在不将计划剂量更改为> 1%的情况下,可以忍受的HU变化程度取决于身体部位和大小,计划算法,治疗束能量和计划类型。软组织HU值的变化比骨骼和空气的HU值变化具有更大的影响。在评估HU变化时,建议使用拟人化体模。关于放射治疗中CT扫描方案优化的已发表工作有限。出版物表明,软组织的HU公差为±20µHU,肺和骨骼的HU公差为±50µHU,这将使治疗计划中的剂量变化限制在<1%。已经审查了与在放射治疗计划中使用CT图像有关的文献,以确定HU改变的可接受水平以及对扫描协议调整图像质量的影响。提出了结论并确定了进一步的工作。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号