首页> 美国卫生研究院文献>Journal of Research of the National Institute of Standards and Technology >A High-Temperature Transient Hot-Wire Thermal Conductivity Apparatus for Fluids
【2h】

A High-Temperature Transient Hot-Wire Thermal Conductivity Apparatus for Fluids

机译:流体的高温瞬态热线导热系数仪

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

A new apparatus for measuring both the thermal conductivity and thermal diffusivity of fluids at temperatures from 220 to 775 K at pressures to 70 MPa is described. The instrument is based on the step-power-forced transient hot-wire technique. Two hot wires are arranged in different arms of a Wheatstone bridge such that the response of the shorter compensating wire is subtracted from the response of the primary wire. Both hot wires are 12.7 µm diameter platinum wire and are simultaneously used as electrical heat sources and as resistance thermometers. A microcomputer controls bridge nulling, applies the power pulse, monitors the bridge response, and stores the results. Performance of the instrument was verified with measurements on liquid toluene as well as argon and nitrogen gas. In particular, new data for the thermal conductivity of liquid toluene near the saturation line, between 298 and 550 K, are presented. These new data can be used to illustrate the importance of radiative heat transfer in transient hot-wire measurements. Thermal conductivity data for liquid toluene, which are corrected for radiation, are reported. The precision of the thermal conductivity data is ± 0.3% and the accuracy is about ±1%. The accuracy of the thermal diffusivity data is about ± 5%. From the measured thermal conductivity and thermal diffusivity, we can calculate the specific heat, Cp, of the fluid, provided that the density is measured, or available through an equation of state.
机译:描述了一种新的设备,该设备用于测量压力为70 MPa,温度为220至775 K时流体的热导率和热扩散率。该仪器基于步进功率强制瞬态热线技术。将两条热线布置在惠斯通电桥的不同臂中,以便从主线的响应中减去较短补偿线的响应。两条热线均为直径为12.7 µm的铂丝,并同时用作电热源和电阻温度计。微型计算机控制电桥调零,施加功率脉冲,监视电桥响应,并存储结果。通过测量液体甲苯以及氩气和氮气来验证仪器的性能。特别是,提出了饱和线附近在298 K和550 K之间的液态甲苯导热系数的新数据。这些新数据可用于说明辐射热传递在瞬态热线测量中的重要性。报告了经过辐射校正的液态甲苯的导热系数数据。导热系数数据的精度为±0.3%,精度约为±1%。热扩散率数据的精度约为±5%。根据测得的导热率和热扩散率,我们可以计算出流体的比热Cp,前提是要测量密度,或者可以通过状态方程获得。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号