首页> 美国卫生研究院文献>Case Reports in Orthopedics >Fracture of a Femoral Revision Stem following a Technical Failure
【2h】

Fracture of a Femoral Revision Stem following a Technical Failure

机译:技术故障后股骨修复干骨折

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

We report about a fracture of a modular, uncemented femoral revision stem (Zimmer, Revitan®) due to a not previously described intraoperative technical problem. During implantation, a small ring, part of the proximal part of the trial stem, was left on the distal part of the definite stem. Following this, the top screwcap of the proximal part of the definite stem could not be tightened properly. However, the stem was thought to be stable, and the situation left. Two and a half years later, the proximal part of the stem fractured and the situation became unstable. It is very useful to know about this ring and that it should always be removed, otherwise, there is possibility that it may be left on the definite distal part of the stem with the possibility of a later fracture.
机译:我们报告了由于先前未描述的术中技术问题而导致的模块化,非骨水泥化股骨翻修柄(Zimmer,Revitan®)的骨折。植入过程中,试验杆近端部分的小环留在确定杆的远端。此后,定柄近端的顶部螺帽无法正确拧紧。但是,人们认为茎是稳定的,情况就此消失了。两年半后,茎的近端骨折,情况变得不稳定。了解此环非常有用,应始终将其取下,否则,可能会将其留在茎的确定的远端部分,并可能随后断裂。