Indirect arterial blood pressure measurement in healthy anesthetized catsusing a device that combines oscillometry with photoplethysmography

机译:健康麻醉猫的间接动脉血压测量使用将示波法与光体积描记法相结合的设备

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

We investigated the basic characteristics of indirect arterial blood pressure (ABP) measurement using a device that combines oscillometry and photoplethysmography in cats. Dobutamine was infused intravenously in four anesthetized cats. Direct ABP was measured by a catheter. Indirect ABP was measured from the left forelimb. Dobutamine significantly elevated both systolic arterial pressure (SAP) and mean arterial pressure (MAP) in a dose-dependent manner. The indirect SAP, MAP and diastolic arterial pressure (DAP) values were closely correlated with the direct ABP values (r=0.88, 0.89 and 0.83, respectively). The mean bias for SAP, MAP and DAP was 3.4, 0.2 and −2.4 mmHg, respectively. The indirect ABP measured by this device may be used to reliably monitor ABP changes in anesthetized cats.
机译:我们调查了使用示波法和光体积描记法结合在猫中的设备间接动脉血压(ABP)测量的基本特征。在四只麻醉的猫中静脉内注射多巴酚丁胺。通过导管测量直接ABP。从左前肢测量间接ABP。多巴酚丁胺以剂量依赖性方式显着升高收缩期动脉压(SAP)和平均动脉压(MAP)。间接SAP,MAP和舒张期动脉压(DAP)值与直接ABP值密切相关(分别为r = 0.88、0.89和0.83)。 SAP,MAP和DAP的平均偏差分别为3.4、0.2和-2.4 mmHg。通过此设备测量的间接ABP可用于可靠地监测麻醉猫的ABP变化。