Data from the analytical performance of the Abaxis Piccolo Xpress point of care analyzer in whole blood serum and plasma

机译:来自Abaxis Piccolo Xpress护理分析仪在全血血清和血浆中的分析性能数据

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

The objective of this study was to examine the analytical performance of 14 comprehensive metabolic panel analytes on the Abaxis Piccolo Xpress® Point of Care analyzer in serum, plasma, and whole blood. A method comparison was performed on all three specimen types intended for use on the Piccolo Xpress®: serum, heparinized plasma, and whole blood. This data is also presented in Murata et al. (2015) . This article includes the actual Bland-Altman bias plots of the difference in results obtained for analytes in the comprehensive metabolic panel from the Abaxis Piccolo Xpress and the comparison instrument, the Ortho Vitros.
机译:这项研究的目的是在Abaxis PiccoloXpress®Point of Care分析仪上检测14种综合代谢组分析物在血清,血浆和全血中的分析性能。对打算在PiccoloXpress®上使用的所有三种样本类型进行了方法比较:血清,肝素化血浆和全血。村田等人也提供了这些数据。 (2015)。本文包括从Abaxis Piccolo Xpress和比较仪器Ortho Vitros在综合代谢面板中获得的分析物结果的实际Bland-Altman偏差图。