首页> 美国卫生研究院文献>Turkish Journal of Anaesthesiology and Reanimation >Sudden Appearance of Water in Flowmeter During Air/Oxygen and Sevoflurane Anaesthesia
【2h】

Sudden Appearance of Water in Flowmeter During Air/Oxygen and Sevoflurane Anaesthesia

机译:空气/氧气和七氟醚麻醉期间流量计中水的突然出现

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

Endotracheal intubation was performed, and a water bubbling sound was heard from the anaesthesia device immediately after the release of gases to administer the O2-air-sevoflurane mixture. The flowmeter on the anaesthesia device was then found to be filled with water. The breakdown of the dryer in the medical air compressor system was determined as the source of the problem, since a greasy fluid mixture was released from the air-wall outlets in all rooms. Consequently, the anaesthesia team should keep in mind that problems as seen in the current case might emerge and should be alert.
机译:进行气管插管,并在释放气体以施用O2-空气-七氟醚混合物后,立即从麻醉装置中听到水起泡的声音。然后发现麻醉设备上的流量计充满水。医用空气压缩机系统中干燥机的故障被确定为问题的根源,因为所有房间的空气壁出口都释放出油腻的液体混合物。因此,麻醉小组应牢记在当前情况下可能会出现问题,并应保持警惕。