首页> 美国卫生研究院文献>Sensors (Basel Switzerland) >Evaluation of Aerosol Electrospray Analysis of Metal-on-Metal Wear Particles from Simulated Total Joint Replacement
【2h】

Evaluation of Aerosol Electrospray Analysis of Metal-on-Metal Wear Particles from Simulated Total Joint Replacement

机译:模拟全接头置换对金属磨损金属颗粒进行气溶胶电喷雾分析的评估

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

Wear is a common cause for aseptic loosening in artificial joints. The purpose of this study was to develop an automated diagnostical method for identification of the number and size distribution of wear debris. For this purpose, metal debris samples were extracted from a hip simulator and then analyzed by the electrospray method combined with a differential mobility analyzer, allowing particle detection ranging from several nanometers up to 1 µm. Wear particles were identified with a characteristic peak at 15 nm. The electrospray setup was successfully used and validated for the first time to characterize wear debris from simulated total joint replacement. The advantages of this diagnostic method are its time- and financial efficiency and its suitability for testing of different materials.
机译:磨损是造成人工关节无菌性松动的常见原因。这项研究的目的是开发一种自动诊断方法,以识别磨损碎片的数量和大小分布。为此,从髋部模拟器中提取金属碎片样品,然后通过电喷雾方法结合差动迁移率分析仪进行分析,从而可以检测范围从几纳米到1 µm的颗粒。确认磨损颗粒在15 nm处具有特征峰。电喷雾装置已成功用于首次验证,可从模拟的全关节置换中表征磨损碎片。这种诊断方法的优点是时间和财务效率高,适合测试不同的材料。