首页> 美国卫生研究院文献>Sensors (Basel Switzerland) >Assessing Inter-Sensor Variability and Sensible Heat Flux Derivation Accuracy for a Large Aperture Scintillometer
【2h】

Assessing Inter-Sensor Variability and Sensible Heat Flux Derivation Accuracy for a Large Aperture Scintillometer

机译:评估大孔径闪烁仪的传感器间变化和感热通量推导精度

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

The accuracy in determining sensible heat flux (H) of three Kipp and Zonen large aperture scintillometers (LAS) was evaluated with reference to an eddy covariance (EC) system over relatively flat and uniform grassland near Timpas (CO, USA). Other tests have revealed inherent variability between Kipp and Zonen LAS units and bias to overestimate H. Average H fluxes were compared between LAS units and between LAS and EC. Despite good correlation, inter-LAS biases in H were found between 6% and 13% in terms of the linear regression slope. Physical misalignment was observed to result in increased scatter and bias between H solutions of a well-aligned and poorly-aligned LAS unit. Comparison of LAS and EC H showed little bias for one LAS unit, while the other two units overestimated EC H by more than 10%. A detector alignment issue may have caused the inter-LAS variability, supported by the observation in this study of differing power requirements between LAS units. It is possible that the LAS physical misalignment may have caused edge-of-beam signal noise as well as vulnerability to signal noise from wind-induced vibrations, both having an impact on the solution of H. In addition, there were some uncertainties in the solutions of H from the LAS and EC instruments, including lack of energy balance closure with the EC unit. However, the results obtained do not show clear evidence of inherent bias for the Kipp and Zonen LAS to overestimate H as found in other studies.
机译:参照涡帕协方差(EC)系统在靠近蒂姆帕斯(Timpas)(美国科罗拉多州)的相对平坦且均匀的草地上评估了三种Kipp和Zonen大孔径闪烁仪(LAS)的显热通量(H)的准确性。其他测试表明,Kipp和Zonen LAS单位之间存在固有的变异性,并且偏向于高估H。比较了LAS单位之间以及LAS和EC之间的平均H通量。尽管具有良好的相关性,但根据线性回归斜率,LAS之间的H偏差在6%和13%之间。观察到物理失准会导致排列良好和排列不正确的LAS单元的H溶液之间的散射和偏差增加。 LAS和EC H的比较显示一个LAS单位几乎没有偏差,而其他两个单位将EC H高估了10%以上。检测器对准问题可能导致了LAS之间的可变性,这一点得到了本研究中LAS单元之间功率要求不同的观察的支持。 LAS物理失准可能会导致光束边缘信号噪声以及风致振动对信号噪声的脆弱性,均会影响H的解。此外,在H值方面存在一些不确定性LAS和EC仪器中H的解决方案,包括EC单元缺乏能量平衡关闭。但是,获得的结果并未显示出其他研究中发现的Kipp和Zonen LAS高估H的固有偏差的明确证据。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号