Performance Evaluation of Calypso 4D Localization and Kilovoltage Image Guidance Systems for Interfraction Motion Management of Prostate Patients

机译:Calypso 4D本地化和千伏图像引导系统对前列腺患者分行运动管理的性能评估

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

Prostate cancer represents a model site for advances in understanding inter- and intrafraction motion for radiotherapy. In this study, we examined the correlation of the electromagnetic transponder system/Calypso 4D Localization System with conventional on-board imaging (OBI) using kilovoltage imaging. Initially using a quality assurance (QA) phantom and subsequently using data of seven patients, the vector distances between Calypso- and OBI-recorded shifts were compared using the t-test. For the 30 phantom measurements, the average differences between the measured Calypso offset and the calculated OBI shift were 0.4 ± 0.4, 0.2 ± 0.3, and 0.4 ± 0.3 mm in the lateral, longitudinal, and vertical directions, respectively (p = 0.73, p = 0.91, and p = 0.99, respectively), and the average difference vector for all sessions was 0.8 ± 0.4 mm. For the 259 patient measurements, the average differences between the measured Calypso offset and the calculated OBI shift were 0.7 ± 0.5, 1.1 ± 0.9, and 1.2 ± 0.9 mm in the lateral, longitudinal, and vertical directions, respectively (p = 0.45, p = 0.28, and p = 0.56, respectively), and the average difference vector for all sessions was 2.1 ± 1.0 mm. Our results demonstrated good correlation between Calypso and OBI. While other studies have explored the issue of Calypso/OBI correlation, our analysis is unique in our use of phantom validation and in our performing the patient analysis on an initial population prior to routine setup using Calypso without OBI. Implications for Calypso's role as a QA tool are discussed.
机译:前列腺癌代表了模型部位的发展,可以进一步了解放射治疗的分节间运动和节内运动。在这项研究中,我们检查了电磁应答器系统/ Calypso 4D定位系统与使用千伏成像的常规车载成像(OBI)的相关性。最初使用质量保证(QA)体模,然后使用7位患者的数据,使用t检验比较Calypso和OBI记录的班次之间的向量距离。对于30个幻像测量,在横向,纵向和垂直方向上,测得的Calypso偏移量与计算出的OBI位移之间的平均差分别为0.4±0.4、0.2±0.3和0.4±0.3 mm(p = 0.73,p分别为0.91和p = 0.99),所有会话的平均差矢量为0.8±0.4 mm。对于259例患者的测量结果,所测量的Calypso偏移量与计算出的OBI位移之间在横向,纵向和垂直方向上的平均差分别为0.7±0.5、1.1±0.9和1.2±0.9 mm(p = 0.45,p分别为0.28和p = 0.56),所有会话的平均差矢量为2.1±1.0 mm。我们的结果表明,Calypso与OBI之间具有良好的相关性。尽管其他研究探索了Calypso / OBI相关性的问题,但我们的分析在使用幻像验证以及在使用不带OBI的Calypso进行常规设置之前对初始人群进行患者分析方面是独一无二的。讨论了对Calypso作为质量检查工具的意义。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号