Evaluation of a Small Animal Irradiation System for Animal Experiments Using EBT3 Model GAFCHROMICTM Film

机译:使用EBT3模型GAFCHROMICTM膜评估用于动物实验的小型动物辐照系统

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

In cancer research, small animal models, for example, mice, rats, or rabbits, facilitate the in-depth study of biological processes and the effects of radiation treatment that can lead to breakthrough discoveries. However, the physical quality of small animal irradiation systems has not been previously evaluated. In this study, we evaluate the quality of a small animal irradiation system using GAFCHROMICTM film and a Tough Water Phantom. The profiles and percentage depth dose curves for several irradiation conditions were measured to evaluate the quality of the irradiation system. The symmetry ratios when the table was rotated were 1.1 (no filter), 1.0 (0.5 mm Al filter), 1.0 (1.0 mm Al filter), 1.1 (2 mm Al filter), and 1.0 (filter consisting of 0.5 mm Al combined with 0.1 mm Cu). The results of measuring the percentage depth dose curve showed that the relative doses were 17.5% (10 mm depth), 12.4% (20 mm depth), 9.5% (30 mm depth), and 7.4% (40 mm filter) with no filters inserted, 78.0% (10 mm depth), 61.1% (20 mm depth), 46.9% (30 mm depth), and 35.3% (40 mm depth) when a 1.0 mm Al filter was inserted, and 94.4% (10 mm depth), 81.7% (20 mm depth), 68.1% (30 mm depth), and 54.7% (40 mm depth) when a filter consisting of 1.0 mm Al combined with 0.2 mm Cu was inserted. These physical assessments seem to be necessary especially in vivo experiments because those increase reliability of data obtained from small animal irradiation systems.
机译:在癌症研究中,小型动物模型(例如小鼠,大鼠或兔子)有助于深入研究生物学过程以及可能导致突破性发现的放射治疗效果。但是,小动物辐照系统的物理质量先前尚未评估。在这项研究中,我们评估了使用GAFCHROMIC TM 胶片和Tough Water Phantom进行小型动物辐照系统的质量。测量了几种辐照条件下的轮廓和深度剂量百分比曲线,以评估辐照系统的质量。旋转工作台时的对称比为1.1(无过滤器),1.0(0.5毫米铝过滤器),1.0(1.0毫米铝过滤器),1.1(2毫米铝过滤器)和1.0(由0.5毫米铝与0.1毫米铜)。深度剂量百分比曲线的测量结果表明,不使用过滤器的相对剂量为17.5%(10毫米深度),12.4%(20毫米深度),9.5%(30毫米深度)和7.4%(40毫米过滤器)插入时,插入1.0 mm Al过滤器时为78.0%(10毫米深度),61.1%(20毫米深度),46.9%(30毫米深度)和35.3%(40毫米深度),以及94.4%(10毫米深度) ),当插入由1.0 mm Al和0.2 mm Cu组成的过滤器时,分别插入81.7%(20 mm深度),68.1%(30 mm深度)和54.7%(40 mm深度)。这些物理评估似乎是必要的,尤其是在体内实验中,因为这样可以提高从小型动物辐照系统获得的数据的可靠性。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号