首页> 美国卫生研究院文献>Computational and Mathematical Methods in Medicine >Extraction of Heart Rate Variability from Smartphone Photoplethysmograms
【2h】

Extraction of Heart Rate Variability from Smartphone Photoplethysmograms

机译:从智能手机体积描记图中提取心率变异性

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

Heart rate variability (HRV) is a useful clinical tool for autonomic function assessment and cardiovascular diseases diagnosis. It is traditionally calculated from a dedicated medical electrocardiograph (ECG). In this paper, we demonstrate that HRV can also be extracted from photoplethysmograms (PPG) obtained by the camera of a smartphone. Sixteen HRV parameters, including time-domain, frequency-domain, and nonlinear parameters, were calculated from PPG captured by a smartphone for 30 healthy subjects and were compared with those derived from ECG. The statistical results showed that 14 parameters (AVNN, SDNN, CV, RMSSD, SDSD, TP, VLF, LF, HF, LF/HF, nLF, nHF, SD1, and SD2) from PPG were highly correlated (r > 0.7, P < 0.001) with those from ECG, and 7 parameters (AVNN, TP, VLF, LF, HF, nLF, and nHF) from PPG were in good agreement with those from ECG within the acceptable limits. In addition, five different algorithms to detect the characteristic points of PPG wave were also investigated: peak point (PP), valley point (VP), maximum first derivative (M1D), maximum second derivative (M2D), and tangent intersection (TI). The results showed that M2D and TI algorithms had the best performance. These results suggest that the smartphone might be used for HRV measurement.
机译:心率变异性(HRV)是用于自主功能评估和心血管疾病诊断的有用临床工具。传统上,它是由专用的医用心电图仪(ECG)计算得出的。在本文中,我们证明了HRV也可以从通过智能手机的相机获得的光电容积描记图(PPG)中提取。通过智能手机捕获的PPG为30名健康受试者计算了16个HRV参数,包括时域,频域和非线性参数,并与心电图进行了比较。统计结果表明,PPG中的14个参数(AVNN,SDNN,CV,RMSSD,SDSD,TP,VLF,LF,HF,LF / HF,nLF,nHF,SD1和SD2)高度相关(r> 0.7,P PPG的7个参数(AVNN,TP,VLF,LF,HF,nLF和nHF)与ECG的参数在可接受的范围内完全吻合(<0.001)。此外,还研究了五种不同的检测PPG波特征点的算法:峰点(PP),谷点(VP),最大一阶导数(M1D),最大二阶导数(M2D)和切线交点(TI) 。结果表明,M2D和TI算法具有最佳性能。这些结果表明,智能手机可能用于HRV测量。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号