Absorbed dose to water was measured with ionisation chambers NE 2561 (#267), NE 2581 (#334), NE 2571 (#1028), using the IAEA standard water phantom. The ionisation chamber was inserted in the water phantom at a reference depth dependent on the type of the radiation quality used. Three radiation qualities were used namely 1.25 MeV gamma ray, 6 MV x-rays and 10 MV x-rays. The values of the absorbed dose to water were determined by the NK- and NX- based methods, i.e with the use of IAEA, HPA, NACP, AAPM, NCRP and ICRU protocols. The aim of this study was to make an intercomparison of the results, by taking the IAEA protocol as a standard. The largest deviation contributed by any of these protocols was recorded for each quality. It was found that AAPM, NCRP and ICRU protocols contributed 0.94% for 1.25 MeV gamma ray, NACP contributed 2.12% for the 6 MV x-rays, and NACP contributed 2.35% for 10 MV x-rays. Since the acceptable limit of deviation set by the IAEA for this absorbed dose work is ± 3%, it is clear that the overall deviations obtained were all satisfactory.
机译:使用IAEA标准水模,在电离室NE 2561(#267),NE 2581(#334),NE 2571(#1028)中测量对水的吸收剂量。根据所使用的辐射质量的类型,将电离室以参考深度插入水模中。使用了三种辐射质量,即1.25 MeV伽马射线,6 MV X射线和10 MV X射线。通过基于NK和NX的方法(即使用IAEA,HPA,NACP,AAPM,NCRP和ICRU协议)确定对水的吸收剂量值。这项研究的目的是通过以国际原子能机构的议定书为标准来对结果进行比较。对于每种质量,记录了这些协议中任何一个所造成的最大偏差。已发现,AAPM,NCRP和ICRU协议对1.25 MeVγ射线贡献0.94%,NACP对6 MV X射线贡献2.12%,而NACP对10 MV X射线贡献2.35%。由于国际原子能机构为此吸收剂量工作设定的可接受的偏差极限为±3%,因此很明显,所获得的总体偏差都令人满意。